45
6. Insert cuvette symbol is shown.
7. Take a sample according to
Blood Collection Capillary Sample
or
Blood Collection
Venous Sample
and Control Material
sections. Place the microcuvette into the cuvette
holder (7a) and start measurement as soon as possible but no later than 1 minute after
filling the microcuvette by gently pushing the cuvette holder to its measuring position
(7b).
8. During measurement the measuring window is shown.