Recommended separation distance between Port-
able and mobile RF communications equipment and
HemoCue Albumin 201 Analyzer
The HemoCue systems are intended for use in an electro-
magnetic environment in which radiated RF disturbances
are controlled. The customer or the user of HemoCue
systems can help prevent electromagnetic interference
by maintaining a minimum distance between portable
and RF communications equipment (transmitters) and
HemoCue systems as recommended below, according
to the maximum output power of the communications
equipment.
Rated maxi-
mum output
power of
transmitter
(W)
Separation distance according to
frequency of transmitter (m)
150 kHz
to 80 MHz
d=1.2
√
P
80 MHz to
800 MHz
d=1.2
√
P
800 MHz
to 2.5 GHz
d=2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at maximum output power not
listed above, the recommended separation distances (d)
in meters (m) can be estimated using the equation ap-
plicable to the frequency of the transmitter, where (P) is
the maximum output power rating of the transmitter in
watts (W) according to transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation dis-
tance for the higher frequencyrange applies.
NOTE 2
These guidelines may not apply in all situ-
ations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and
people.
Summary of Contents for Albumin 201
Page 1: ...HemoCue Albumin 201 Operating Manual ...
Page 4: ...4 Components 1 2 4 3 5 6 ...
Page 6: ...6 Start up 1 2a 3 4 2b ...
Page 8: ...8 Set up Printer function 1 3 ...
Page 10: ...10 Specimen collection U Albumin Spot urine morning urine ...
Page 12: ...12 Measurement 1 2 ...
Page 14: ...14 Measurement 3a 3b 3c 4 5 ...
Page 16: ...6b 16 Measurement 6a 6c ...
Page 18: ...18 Maintenance 1 2 3 4 ...
Page 20: ...Turbidity scale 20 ...
Page 35: ......