US
15
14
3.
Open the lid and place the filled
cuvette into the cuvette holder
(3a). It is important that the
cuvette “snaps” properly into
the cuvette holder (3b). Start
measurement as soon as possible
but no later than 30 seconds
after filling the microcuvette by
gently closing the lid (3c).
4.
During measurement the hour
glass
6
is displayed with three
fixed dashes.
5. Within 90 seconds, the result is
displayed in mg/L. The result will
remain on the display until the
lid is opened.
Measurements below 7 mg/L are
displayed as “LLL”.
Measurements above 150 mg/L
are displayed as “HHH”.
If the display shows an error
code, an error has occurred.
See the Troubleshooting Guide.
Do not remeasure the cuvette.
Note that the measurement is
not completed until the steps
in paragraph 8 have been per-
formed and the display shows
three flashing dashes.