19
18
The cuvette holder should be cleaned
after each day of use. The cover
glasses of the optical unit should be
cleaned when directed to do so in the
Troubleshooting Guide.
Make sure that the analyzer is switched
off and that the display is blank.
Open the lid.
1. The cuvette holder is released by
pressing the area marked with
arrows using the device included
with the system.
2. Press the tab downwards and
draw the cuvette holder towards
you until it becomes detached.
Use a mild detergent or alcohol to
clean the cuvette holder.
3. The cover glasses of the optical
unit, both beneath and above the
cuvette holder area, may be cleaned
using the HemoCue Cleaner*. If a
HemoCue Cleaner is not available,
a cotton tipped swab moistured
with water or alcohol without
additives may be used.
4. Wait 15 minutes before replac-
ing the cuvette holder.
Use alcohol or mild soap deter-
gent for cleaning the outer sur-
face of the instrument.
* One HemoCue Cleaner is included
with the analyzer. Contact HemoCue
America to order more.
US