Symptom
Explanation
Action
E08
The absorbance is too high.
1. Light blocking item in the cuvette holder.
1a. Check that the instrument and cuvettes are being
used according to the HemoCue Albumin 201 system
Operating Manual and instructions for use.
1b. The instrument needs service. Contact HemoCue America.
E09
1. This is caused by high absorbance-values due to
immunoinhibition.
1. The albumin concentration is too high to be quantified
by the HemoCue Albumin 201 Analyzer.
E10
This can happen during use and is an error that pre-
vents further measurements. A read or write opera-
tion to the EEPROM did not succeed.
1. The EEPROM is out of order.
1a. Turn off the instrument and turn it on again
after 30 seconds. Take a new cuvette and repeat the
measurement.
1b. The instrument needs service. Contact HemoCue America.
E13
1. This can be seen during use and is an error that
prevents further measurements. A read or
write operation to the RAM did not succeed.
1. The instrument needs service. Contact HemoCue America.
E21
1. Testing a configuration area in EEPROM returned
as an erroneous checksum.
1. Turn off the instrument and turn it on again after 30
seconds. If the problem continues, the instrument
needs service. Contact HemoCue America.
E23
Data Error Real Time Clock.
Real Time Clock backup battery has been drained.
The backup battery needs to be replaced. Contact HemoCue
America.
E25
Can be seen when starting the instrument and is
an error that prevents further measurements. The
checksum of the calibration area is not valid.
1. The instrument needs to be calibrated.
1. The instrument needs service. Contact HemoCue America.
E30
1. Fault or dirt in the optic or electronic units.
1a. Turn off the analyzer and clean the optic unit using the
HemoCue Cleaner.
1b. The analyzer needs service. Contact HemoCue America.
23
22
Summary of Contents for Albumin 201
Page 1: ...HemoCue Albumin 201 Operating Manual ...
Page 4: ...4 Components 1 2 4 3 5 6 ...
Page 6: ...6 Start up 1 2a 3 4 2b ...
Page 8: ...8 Set up Printer function 1 3 ...
Page 10: ...10 Specimen collection U Albumin Spot urine morning urine ...
Page 12: ...12 Measurement 1 2 ...
Page 14: ...14 Measurement 3a 3b 3c 4 5 ...
Page 16: ...6b 16 Measurement 6a 6c ...
Page 18: ...18 Maintenance 1 2 3 4 ...
Page 20: ...Turbidity scale 20 ...
Page 35: ......