Analog Module Operation
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
29
Be
for
e
You Beg
in
In
stallation
A
na
log O
per
atio
n
R
S–48
5 Op
er
ation
M
ain
te
n
a
n
ce
Figure 3-10 Example test point voltage: 0.64 V equals 40 mTorr of N
2
or air
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1
2
5
10
20
50
100
Test Point Voltage
Pressure in mT
orr for N
2
or air
0.64 V
40
mTorr
Summary of Contents for GRANVILLE-PHILLIPS Convectron ATM