Chapter 3
16
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
Before putting the module into operation, you must perform the following
procedures:
1.
Install the module in accordance with the instructions on pages 5-13.
2.
Develop a logic diagram of the process control function.
3.
Use Table 3-2 to record the proposed setpoint in volts, corresponding
pressure setting in Torr, and activation direction for each relay.
Table 3-1
Front panel features
Feature
Description
Measure at:
Set to:
CONVECTRON ATM
ADJUST potentiometer
Enables Convectron gauge calibration at
atmospheric pressure
CONVECTRON
ANALOG and
COMMON test points
5.534 V at
760 Torr of N
2
CONVECTRON VAC
ADJUST potentiometer
Enables Convectron gauge calibration at
vacuum chamber pressure
CONVECTRON
ANALOG and
COMMON test points
0.375 V at
1 x 10
–4
Torr
of N
2
DIAPHRAGM ATM
ZERO adjust
potentiometer
Enables calibration of differential pressure
sensor zero
DIAPHRAGM ZERO and
COMMON test points
4.000 V
SP1 ADJ potentiometer
Enables setting of vacuum pressure at which
relay 1 activates and deactivates
SETPOINT1 and
COMMON test points
Application
specific
SP2 ADJ potentiometer
Enables setting of differential pressure at which
relay 2 activates and deactivates
SETPOINT2 and
COMMON test points
Application
specific
Test points
Enable checking of values for:
• Setpoint 1 voltage
• Setpoint 2 voltage
• Convectron gauge calibration at
atmospheric pressure
• Convectron gauge calibration at vacuum
chamber pressure
• Differential pressure sensor zero calibration
See cells above in this
column
Application
specific
Setpoint indicators
• SETPOINT1 is green when relay 1 is
activated
• SETPOINT2 is green when relay 2 is
activated
Not applicable (used for
monitoring relays)
Not applicable
Power ON indicator
Is green when module power supply is ON
Not applicable (used for
monitoring module
power supply)
Not applicable
Summary of Contents for GRANVILLE-PHILLIPS Convectron ATM
Page 2: ......
Page 4: ......
Page 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Page 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Page 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Page 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Page 97: ......