Setpoint 2: differential
pressure sensor
Refer to Figure 3-12 and use a a high-precision, high-input impedance
(Z
in
> 1 M
Ω
) DVM to adjust relay 2.
1.
Use the following equation or the data in Figure 3-12 to determine the
analog output voltage that corresponds to the differential pressure, in
Torr, at which relay 2 will activate and deactivate:
Figure 3-12 Setpoint relay 2 (differential pressure sensor) analog voltage
Output voltage
4.000
P
differential
0.004 V
×
+
=
-1000
-750
-500
-250
0
250
0
1.0
2.0
3.0
4.0
5.0
6.0
Differential Pressure (Torr)
Setpoint 2 testpoint v
oltage