Chapter 3
26
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
3.3
Setpoint relays
The module includes two single-pole, double throw (SPDT) relays. Each
relay has a programmable setpoint. The setpoint is a voltage level that
corresponds to a specified pressure at which the relay activates and
deactivates. See Table 3-6.
In default mode, relay 1 activates with decreasing vacuum chamber
pressure and deactivates at a higher pressure than the activation pressure,
as illustrated in Figure 3-7.
In default mode, relay 2 activates with increasing differential pressure and
deactivates at a lower pressure than the activation pressure, as illustrated in
Figure 3-8 on page 27. Use internal jumpers to reconfigure relay 2
activation direction. See page 33.
Figure 3-7
Setpoint relay 1 behavior
Table 3-6
Setpoint relay activation direction, range, and hysteresis
Relay
Assignment
Activation direction
Range
Hysteresis
Setpoint 1
Vacuum chamber pressure
Decreasing pressure
1.0 x 10
–4
to 9.99 x 10
2
Torr
30 mV
Setpoint 2
Differential pressure
Increasing pressure
–750 to +125 Torr
8 Torr
Deactivate
Activate
Deactivated
Activated
Relay activated
30mV
hysteresis
Time
Vacu
um chamber pressure
Summary of Contents for GRANVILLE-PHILLIPS Convectron ATM
Page 2: ......
Page 4: ......
Page 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Page 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Page 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Page 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Page 97: ......