Heidolph 506-21100-00 Operating Instructions Manual Download Page 54

55

Attachements

Technical data

General device data

Model

Magnetic stirrer 

Hei-PLATE Mix

'

n

'

Heat Core

+

Dimensions (W × H × D)

168 × 101 × 299 mm

Usable surface heating plate

 135 mm or 

 145 mm 

(depending on model, see ordering data)

Weight

approx. 3 kg

Maximum permissible load

25 kg

Drive

EC motor, left-turning

Speed range

100 – 1,400 rpm

Adjustment accuracy

5 rpm

Heating power

800 W at 230 V (EU)
600 W at 115 V (US)

Heating temperature range

20 – 300 °C

Heating control

PID

Adjustment accuracy

1 K

Measurement accuracy (DIN 

IEC 751 Class a)

±0.2 K, plus tolerance PT1000

Measurement resolution

1 K

Display

LCD display

Protection class (EN 60529)

IP42

Acoustic pressure

< 50 db(A)

Attachements

Summary of Contents for 506-21100-00

Page 1: ...Betriebsanleitung Operating instructions Hei PLATE Mix n Heat Core Safe Heating and Mixing...

Page 2: ...Originalbetriebsanleitung Seite 6 30 Translation of the original instructions Page 36 64...

Page 3: ...erung 8 Akklimatisierung 8 Umgebungsbedingungen 8 K lteanwendung 8 Sicherheit Allgemeine Sicherheitshinweise 9 Elektrische Sicherheit 9 Betriebssicherheit 9 Arbeitssicherheit 10 Pers nliche Schutzausr...

Page 4: ...uf 16 Funktion Auto Standby 16 Sicherheitsfunktion 17 Sensorkalibrierung 17 Bedienung Allgemeine Hinweise 18 Dr ck Dreh Regler 19 Werte anpassen 19 Heiz Mix Funktion ein ausschalten 19 Sollwerte anzei...

Page 5: ...inigungshinweise 27 Reparaturen 27 Wartung 28 Entsorgung 28 Kontaktdaten Deutschland sterreich Schweiz 29 Garantieerkl rung 29 Unbedenklichkeitserkl rung 30 Zertifikate EU Konformit tserkl rung 61 UKC...

Page 6: ...re Verletzungen Verbotszeichen weisen auf Handlungsweisen oder Situationen hin die zu unterlassen bzw zu vermeiden sind Bei Nichtbeachtung drohen Personen und oder Sachsch den Gebotszeichen weisen auf...

Page 7: ...zum Konsum durch Menschen oder Tiere oder zur Anwendung am Menschen oder Tier bestimmt sind ausschlie lich in analytischen Prozessen oder unter labor hnlichen Bedingungen zul ssig Vern nftigerweise vo...

Page 8: ...limatisieren um eventuellen Sch den durch Betauung oder Kondensation vorzubeugen Verl ngern Sie die Akklimatisierungsphase ggf bei sehr hohen Temperaturunterschieden Umgebungsbedingungen Das Ger t dar...

Page 9: ...ber eine Fehlerstrom Schutzeinrichtung RCD abgesichert ist Betreiben Sie das Ger t ausschlie lich mit der mitgelieferten Netzanschlussleitung Stellen Sie vor jeder Inbetriebnahme sicher dass weder das...

Page 10: ...e PSA ist abh ngig vom jeweiligen Einsatzbereich und von den einge setzten Medien und Chemikalien vom Betreiber festzulegen und bereitzustellen Die entsprechende Unterweisung des Personals liegt aussc...

Page 11: ...Control Points Das HACCP umfasst die folgenden Kriterien Gefahrenanalyse Ermittlung kritischer Kontrollpunkte Festlegung kritischer Grenzwerte Etablierung eines Systems zur berwachung und Kontrolle d...

Page 12: ...n sowie besondere rtliche Bestimmungen zwingend zu beachten Im Falle von Zuwiderhandlungen erlischt jeglicher Garantieanspruch gegen ber Heidolph Instruments F r s mtliche Sch den die aus eigenm chtig...

Page 13: ...13 Ger tebeschreibung Ger tebeschreibung Mechanischer Aufbau...

Page 14: ...tabstand von 10 cm zwischen dem Magnetr hrer und anderen Ger ten und Strukturen sicher Stromversorgung Betreiben Sie das Ger t ausschlie lich mit der mitgelieferten Netzanschlussleitung Beachten Sie d...

Page 15: ...Temperatursensor PT1000 Sicherheitsfunktion aktivieren deaktivieren Men Ger teeinstellungen ffnen Gehen Sie wie folgt vor um das Men Ger teeinstellungen zu ffnen Schalten Sie das Ger t aus Tasten LED...

Page 16: ...AR auf on setzen um die Funktion zu aktivieren Parameter AR auf off setzen um die Funktion zu deaktivieren Sanftanlauf Bei aktivierter Funktion Sanftanlauf wird die Drehzahl nach dem Einschalten der...

Page 17: ...auf off setzen um die Funktion zu deaktivieren Parameter SAF auf on setzen um die Funktion zu aktivieren Sensorkalibrierung Nur relevant und verf gbar im Betrieb mit externem Temperatursensor siehe a...

Page 18: ...gleichm ige Verteilung der Beh lter auf der Heizplatte Achten Sie insbesondere auf eine gleichm ige Verteilung der Gewichtslast wenn unterschiedlich gro e und oder unter schiedlich bef llte Gef e gle...

Page 19: ...n Dr ck Dreh Regler Rotation um die Mix Funktion ein und auszuschalten Sollwerte anzeigen Bewegen Sie die beiden Regler im Normalbetrieb um einen Rastpunkt im oder gegen den Uhrzeigersinn um den einge...

Page 20: ...tur einzustellen Einstellbereich 20 300 C Der Sollwert kann im laufenden Betrieb jederzeit angepasst werden Dr cken Sie auf den Dr ck Dreh Regler Temperatur um die Heizfunktion einzuschalten Bei einge...

Page 21: ...1 400 rpm Dr cken Sie auf den Dr ck Dreh Regler Rotation um die Mix Funktion einzuschalten Bei eingeschalteter Mix Funktion leuchtet der LED Ring des Dr ck Dreh Reglers Rotation wei Auf dem Display wi...

Page 22: ...oben ab einer Temperatur von 50 C stets die erforderliche pers nliche Schutzausr stung hitzebest ndige Handschuhe Augenschutz Sicherheitskleidung Vorgehensweise Stellen Sie sicher dass der externe Tem...

Page 23: ...ine Heizfunktion Sollwert Heiztemperatur liegt unterhalb der Ist Temperatur Sollwert anpassen Heizplattendefekt technischen Service kontaktieren Temperaturanzeige bei angeschlossenem Temperatursensor...

Page 24: ...eizplatte Tsoll 25 C E21 Externer Temperatursensor Medienber hrung im laufenden Prozess verloren E22 Externer Temperatursensor ohne Medienber hrung beim Einschalten E23 Sicherheitstemperatur extern be...

Page 25: ...3 kg Maximal zul ssige Belastung 25 kg Antrieb EC Motor linksdrehend Drehzahlbereich 100 1 400 U min Einstellgenauigkeit 5 U min Heizleistung 800 W bei 230 V EU 600 W bei 115 V US Heiztemperaturbereic...

Page 26: ...i bis zu 50 rel Luftfeuchte linear abnehmend Aufstellh he bis 2 000 ber NN Lieferumfang Komponente Menge Produktnr Hei PLATE Mix n Heat Core 1 506 21100 00 Bedienungsanleitung 1 01 005 006 76 Garantie...

Page 27: ...VORSICHT Sch den am Ger t Bei unsachgem er Reinigung besteht die Gefahr die Oberfl chen des Ger ts zu besch digen Durch eindringende Fl ssigkeit k nnen die elektronischen Bauteile im Inneren des Ger...

Page 28: ...gung Beachten Sie bei der Entsorgung des Ger ts die Bestimmungen der WEEE Richtlinie 2012 19 EU sowie deren Umsetzung in nationales Recht im Anwenderland Beachten Sie bei der Entsorgung von Ger tebatt...

Page 29: ...vom Garantieanspruch sind Glas und Verschlei teile Transportsch den sowie Sch den die auf einen unsachgem en Umgang oder eine nicht bestimmungsgem e Verwendung des Produkts zur ckzuf hren sind Der Ga...

Page 30: ...d oder die Umwelt aus Bitte ankreuzen und Angaben erg nzen NEIN JA Wenn JA mit welchen Substanzen kam das Ger t in Ber hrung Wenn NEIN welche Ma nahmen zur Reinigung und oder Dekontamination wurden du...

Page 31: ...Translation of the original instructions Page 36 65...

Page 32: ...8 Acclimatization 38 Permissible ambient conditions 38 Processing liquids at extreme cold temperatures 38 Safety General safety information 39 Electrical safety 39 Operational safety 39 Work safety 40...

Page 33: ...function 46 Safety function 47 Sensor calibration 47 Operation General notes 48 Push and rotary controls 49 Adjust values 49 Switch heating mix function on off 49 Display set values 49 Lock function 4...

Page 34: ...ng instructions 57 Repairs 57 Maintenance 58 Disposal 58 Contact information Heidolph international 59 Warranty Statement 59 Declaration of no objection 60 Certifications EU Declaration of Conformity...

Page 35: ...te injury or material environmental damage Prohibition signs indicate actions or situations that are to be omitted or avoided Failure to comply may result in personal injury and or damage to property...

Page 36: ...in this manual is not considered as intended Due to its design the device in its delivery condition may only be used in analytical processes or in laboratory like conditions in the food cosmetics and...

Page 37: ...f use If necessary extend the acclimatization phase if the temperature differences are very high Make all supply connections power supply tubing only after the device has been acclimatized Permissible...

Page 38: ...supplied power supply cord provided with the device Prior to use check that the device and the power supply cord are free of visible damage Have repairs and or maintenance work on the device carried o...

Page 39: ...adequate safety distance Do not store objects in the working and hazardous area of the device during operation Personal protective equipment PPE The operator must determine and provide the necessary...

Page 40: ...ication of critical control points Definition of critical limit values Establishment of a system for monitoring and controlling critical hazard control points CCP Corrective actions for uncontrollable...

Page 41: ...ring and occupational medicine as well as particular local regulations Noncompliance will invalidate any warranty against Heidolph Instruments The operator is solely liable for all damage resulting fr...

Page 42: ...43 Device description Device description Mechanical design...

Page 43: ...distance of 10 cm between magnetic stirrers and other devices and structures devices with heating function only Power Supply Always use the supplied power supply cord provided with the device Observe...

Page 44: ...tart function Calibration of external temperature sensor PT1000 Activating deactivating the safety function Open the settings menu To open the settings menu proceed as follows Turn off the device on o...

Page 45: ...pen the settings menu page 45 Set AR to on to activate the function Set AR to off to disable the function Soft start When the soft start function is activated the speed is gradually increased to the s...

Page 46: ...sable the function Set SAF to on to activate the function Sensor calibration Only relevant and available operating the device with an external temperature sensor see also section Operation with extern...

Page 47: ...d fall Always ensure that the vessels are evenly distributed on the shaking platform Pay particular attention to an even distribution of the weight load when vessels of different sizes and or differen...

Page 48: ...on on and off Press the rotation control to turn the mixing function on and off Display set values During normal operation move the two controls one detent position clockwise or count er clockwise to...

Page 49: ...ing range 20 300 C The setpoint can be adjusted at any time during operation Press the temperature control to activate the heating function When the heating function is activated the LED ring of the t...

Page 50: ...nction When the mixing function is activated the LED ring of the rotation control lights up White The displayed rotation speed value continuously updates until the setpoint is reached The setpoint can...

Page 51: ...use the necessary personal protective equipment heat resistant gloves eye protection safety clothing to process samples from a temperature of 50 C Procedure Ensure that the external temperature sensor...

Page 52: ...ed set to zero increase rotation speed No heating function Heating temperature setpoint below current temperature adjust setpoint Heating plate defective contact technical service Temperature display...

Page 53: ...Thotplate Tset 25 C E21 External temperature sensor media contact lost in the current process E22 External temperature sensor without media contact when switching on E23 External safety temperature ex...

Page 54: ...kg Maximum permissible load 25 kg Drive EC motor left turning Speed range 100 1 400 rpm Adjustment accuracy 5 rpm Heating power 800 W at 230 V EU 600 W at 115 V US Heating temperature range 20 300 C...

Page 55: ...midity 32 C 40 C up to 50 rel humidity decreasing linearly Maximum height above sea level 2 000 m Scope of delivery Item Quantity Product no Hei PLATE Mix n Heat Core 1 506 21100 00 Operating instruct...

Page 56: ...istent contamination can be removed with mild soapy water CAUTION Damage to the device Improper cleaning can damage the surfaces of the device Penetrating liquid can damage the electronic components i...

Page 57: ...visions of the WEEE Directive 2012 19 EU and its transposition into national law in the country of use When disposing of portable batteries observe the provisions of the European Battery Directive 201...

Page 58: ...against mate rial and manufacturing defects Excluded from the warranty are glass and wear parts transport damage and damage resulting from improper handling or non in tended use of the product For reg...

Page 59: ...lth the envi ronment and or biohazardous Please mark with a cross and complete the information NO YES If YES with which substances did the device come into contact If NO what cleaning and or decontami...

Page 60: ...is in compliance with the basic requirements of all applicable EU directives stated below with regard to design type of model sold and manufactured by us This certificate will be invalid if the produ...

Page 61: ...eat Expert 506 3xxx0 xx Hei PLATE Mix n Heat Ultimate 506 4xxx0 xx Hei PLATE Mix 20 l 506 5xxx0 xx Radleys TECH 506 2xxx3 xx Description Magnetid stirrer with or without heating plate and human machin...

Page 62: ...63 RoHS Konformit tserkl rung RoHS Declaration of Conformity Zertifikate Certifications...

Page 63: ...g which the hazardous substances contained in electrical and electronic products will not leak or mutate under normal operating conditions During normal use by the user such electrical and electronic...

Page 64: ...Pb Hg Cd Cr VI PBB PBDE Apart from the disclosures in the above table the subassemblies are not intentionally manufactured or formulated with lead Pb mercury Hg cadmium Cd hexavalent chromium CrVI pol...

Page 65: ...ument unterliegt in gedruckter Form keinem nderungsdienst der jeweils neueste Ausgabestand steht auf unserer Homepage zum Download zur Verf gung Subject to change without notice The printed version of...

Reviews: