Commissioning
External temperature sensor
The port used to connect the external temperature sensor (PT1000) is located under
the protective cover on the rear of the device (see figure above).
➜
Lift the protective cover for connecting the sensor cable by hand. No tools are
required to open the protective cover.
➜
Make sure that the connector of the sensor cable is properly connected to the
device and locked.
Switching the device on/off
To turn on the device, press the on/off button. When the device is switched on, the on/
off button lights up green.
To turn off the device, press the on/off button again. When the device is turned off, the
on/off button lights up white.
Settings
The following adjustments can be made in the device settings:
–
Enable/disable auto standby
–
Activating/deactivating the [automatic restart] function
–
Activating/deactivating the [soft start] function
–
Calibration of external temperature sensor (PT1000)
–
Activating/deactivating the safety function
Open the settings menu
To open the
settings menu
, proceed as follows:
➜
Turn off the device (on/off button lights up white).
➜
Press simultaneously and hold both the push-and-rotary controls
[temperature]
and
[rotation]
.
➜
Press the on/off button until
AR off
(first setting option) appears on the display.
➜
Release the push-and-rotary controls and the on/off button.
➜
Use the
[temperature]
control to select one after another the following options:
–
AR
: Activate or deactivate automatic restart
–
SoS
: Activate or deactivate soft start
–
ASd
: Activate or deactivate auto standby
–
SAF
: Activate or deactivate the security function
–
CAL
: Calibrate the connected temperature sensor
➜
Use the
[rotation]
control to select the desired mode:
–
[
on
] or [
off
] for automatic restart, soft start, auto standby and safety function
–
Temperature deviation in
°C
for sensor calibration
If no selection is made within about five seconds, the menu is auto
-
matically closed and the device is in the on state!
Summary of Contents for 506-21100-00
Page 1: ...Betriebsanleitung Operating instructions Hei PLATE Mix n Heat Core Safe Heating and Mixing...
Page 2: ...Originalbetriebsanleitung Seite 6 30 Translation of the original instructions Page 36 64...
Page 13: ...13 Ger tebeschreibung Ger tebeschreibung Mechanischer Aufbau...
Page 31: ...Translation of the original instructions Page 36 65...
Page 42: ...43 Device description Device description Mechanical design...
Page 62: ...63 RoHS Konformit tserkl rung RoHS Declaration of Conformity Zertifikate Certifications...