Technical Specifications
Angle
with lens for retina examination
with lens for anterior segment
with lens for optic nerve head
examination
50° (30° x 40°)
12 x 16 mm
25° (20° x 15°)
Refraction compensation
±
15 dpt
Distance to cornea
50°/25°
lens
lens for anterior eye segment
10 mm
117 mm
Filter:
Exciter and barrier filter (optional) for fluorescein angiography
Red free filter
Neutral density filter T=25%
Halogen Lamp
6V / 35W with justified socket
item order no. 18947
Safety Analysis for the HRC Light source
in accordance to ISO 15004 2:2007
“Opthalmic instruments – Fundamental
requirements and test methods”
The light emitted from the HRC is potentially
hazardous. The longer the duration of
exposure, the greater the risk f ocular damage.
Exposure to light from this instrument when
operated at maximm intensity will exceed the
safety guideline after 30 minutes.
Video system:
Color
camera
Black and white camera
PAL/CCIR 768 x 576 pixel, 25 fps
PAL/CCIR 768 x 576 pixel, 25 fps
(fps - frames per second)
Mechanical data:
HRC:
height:
width
length
weight
Power supply unit:
height:
width
depth
weight
approx. 120 mm
approx. 70 mm
approx. 150 mm
approx. 1.8 kg
approx. 67 mm
approx. 281 mm
approx. 220 mm
approx. 2.5 kg
Electrical data:
Input
Voltage
fuse
Input
Voltage
fuse
110V – 115 V
2 x T 2.0 AL
220V – 240 V
2 x T 2.0 AL
HRC Installation Instructions Vers. 001