Software Settings
The COM port setting for the footswitch connection in the hrc.ini file has to match the
COM port indicated in the Windows Device Manager. Otherwise the HEYEX software will
not find the footswitch when opening the acquisition window.
The COM port default setting in the hrc.ini file is COM1.
Please check the COM port setting in the device manager upon installation. If it does not
match the setting in the hrc.ini file, change the setting in the hrc.ini accordingly (see red
circles in figure 3 and 4 below).
The USB-serial adapter has to stay connected to the same USB port for the system to
operate.
Figure 2:
My computer > right click > properties (=System Properties) > Hardware > Device Manager