Healy VP1000-5 Troubleshooting Manual Download Page 14

14

Appendix C

VP1000 Vane & Rotor Service & Replacement Guide

VP1000 Vane & Rotor Service & Replacement Guide

Caution

Disconnect power before beginning service.

1.  The work area 

must

be clean and have suf cient lighting.

2.  Disconnect the vapor piping connected to the 

IN

 and 

OUT

 ports of the VP1000 cover assembly.

3.  Remove the four Allen head screws and lock washers that secure the pump cover assembly to the pump housing and 

remove the cover carefully.

Caution

Use a spill cloth when removing the cover, as there may be some gasoline inside the pump cavity

.

4.  Carefully turn the rotor assembly by hand until the shaft key notch 

is at the 12 o’clock position. (See Figure 1)

5.  Remove the rotor, vanes and shaft key from the pump housing.

Note:

Place your hand or a container under the rotor while removing. 

Do not use any sharp objects that would scratch the surfaces of 

the pump cavity, pump shaft, rotor, or vanes.

6.  Rotate the shaft by hand. If the shaft does not rotate freely, the 

entire vacuum pump needs replacement (p/n VP1000-5).

7.  If the rotor and vanes are cracked, chipped, excessively worn or 

excessively dirty, the rotor and vanes should be replaced because 

cleaning will not remedy these conditions (p/n VP1000VRC or 

VP1000VRC-P).

8.  If there is no visible damage, use a lint-free cloth with isopropyl 

alcohol to clean the rotor and vanes.

9.  Using a lint-free cloth with isopropyl alcohol, thoroughly clean: the 

inside of the pump ring and rear of the pump cavity, the rotor shaft, 

and the inside of the pump cover.

10.  Reposition the shaft (if necessary) so that the shaft key notch is 

in the 12 o’clock position. Install the cleaned original or new shaft 

key onto the shaft.

11.  Carefully install the cleaned original or new rotor onto the shaft 

followed by the cleaned original or new vanes into the rotor.

Note:

The rotor assembly should slide on to the shaft easily, without 

excessive force. (Rotors and vanes are reversible)

Figure 1

VP1000VRC

Carbon Rotor

(limited availability)

Shaft Key

Notch

Vane

Shaft Key

Notch

Vane

Figure 2: 

VP1000VRC-P

Enhanced

Carbon Rotor

12.  Lightly lubricate and install the new O-Ring for the pump housing.

Note:

Do not allow any lubricant to get inside the pump housing.

13.  Install the pump cover using the four Allen head screws and lock washers removed in step 3 and cross tighten.

Note:

Use caution when sliding the pump cover over the O-Ring seal to prevent cutting or tearing.

14.  Re-connect the vapor piping to the 

IN

 and 

OUT

 ports of the pump cover assembly that was removed in Step 2.

15.  Re-apply power. Test for normal operation. (See VP1000 Vacuum Performance Test Procedure)

Note:  See CARB Executive Order VR-201 for complete installation, operation, and maintenance (IOM) details.

Summary of Contents for VP1000-5

Page 1: ...anklin Fueling Systems 3760 Marsh Rd Madison WI 53718 USA Tel 1 608 838 8786 800 225 9787 Fax 1 608 838 6433 www franklinfueling com Manual Revision Date Changes from Previous Revision 405291001 2 Mar...

Page 2: ...Under Current Fault Troubleshooting 8 Section Four Over Current Fault Troubleshooting 9 Section Five VP1000 5 Performance Testing 10 Appendix A VP1000 5 Installation and Testing 12 Appendix B Side B...

Page 3: ...environment or by using any other reasonable means available to ensure the safety of service personnel Follow all federal state and local laws governing the installation of this equipment When no oth...

Page 4: ...202 A Healy Phase II System outside California in North America includes one VP1000 5 vacuum pump per dispenser and the appropriate Healy hanging hardware including Model 800 nozzles at each fueling...

Page 5: ...nstalled per dispenser to accommodate the vapor recovery needs for both sides of the dispenser When the dispenser is authorized the MC100 1 will activate the VP1000 5 simultaneously and run at a low s...

Page 6: ...oduct The MC100 1 s solenoid disconnect relay provides interruption of the solenoid valve neutral circuit when the VP1000 5 is in a fault state Figure 2 describes the meanings of the indicator lights...

Page 7: ...Performance Testing in Section 5 Is a fault condition present on the MC100 1 Reset the system by removing the white reset plug on the MC100 1 for a minimum of 20 seconds Re insert the white reset plu...

Page 8: ...er VR 201 No Verify the performance of the VP1000 5 per B 5 of the A B sheet Refer to Appendix B Yes Is the vacuum level produced by the VP1000 5 between 60 90 WC Disconnect the white reset plug on th...

Page 9: ...the VP1000 5 between 60 90 WC Disconnect the white reset plug on the MC100 1 Remove the VP1000 5 front cover to evaluate the rotor vanes for wear and or damage Refer to Appendix C Check for debris and...

Page 10: ...mance of the VP1000 5 per B 5 of the A B sheet Refer to Appendix B Is the vacuum level produced by the VP1000 5 between 60 and 90 WC Disconnect the white reset plug on the MC100 1 Remove the VP1000 5...

Page 11: ...e VP1000 5 increase from low to high speed when both sides of the dispenser are activated Verify the wiring of the MC100 1 Refer to Appendix E No Retest Continued from previous page Remove the vacuum...

Page 12: ...ow shows a vacuum gauge face with the acceptable vacuum level highlighted when measured with a VP1000 5 vacuum pump The acceptable vacuum level from the VP1000 5 is between 60 and 90 WC Readings of 3...

Page 13: ...est until all leaks have been repaired 8 Record data in Section B 4 Note If the initial vacuum reading is less than 60 wc it could indicate a problem with the VP1000 Remove the dispenser from service...

Page 14: ...e damage use a lint free cloth with isopropyl alcohol to clean the rotor and vanes 9 Using a lint free cloth with isopropyl alcohol thoroughly clean the inside of the pump ring and rear of the pump ca...

Page 15: ...activate Note For unihose dispensers conduct individual tests for each product grade on each side of the dispenser to ensure that the same LED activates for all grades on the same side If the other L...

Page 16: ...C should be measured between either red and orange wires When no fault is present 5VDC should be measured between purple and orange wires When fault is present 0VDC should be measured between purple a...

Page 17: ...ACH and JUMPER terminals to create a short 3 With the jumper wire in place re insert the white reset plug The yellow power light should now be illuminated 4 Authorize a nozzle on Side A of the dispens...

Page 18: ...P1000 MC100 interface check one Red Light is FLASHING over current Red Light is SOLID under current Red Light is OFF Dispenser Integrity inches of water column Starting Vacuum ________ Ending Vacuum _...

Page 19: ...19 2012 FFS 405291001 Rev2...

Reviews: