124
Connection
HBM
QuantumX
Retrofilling the TEDS chip in the transducer plug when using the Smart module:
The Smart module cannot read TEDS data directly saved to the IEPE transducer. The
QuantumX plug can be retrofitted with TEDS to enable the Smart module to be read and to
automate channel configuration according to the IEPE transducer.
TEDS Editor enables settings to be made.
S
TEDS
−
specific settings:
f
”High Level Voltage Output Sensor”
f
Physical Measurand: Acceleration (m2/s or g)
f
Electrical range: st/
−
10 V
f
Power requirements: required
S
Complete the data sheet according to the transducer.
S
Excitation level: 24 V nominal
CAUTION
Replacing the IEPE transducer on the Smart module may result in incorrect device
settings.