Structure and mode of operation
20
A4359-1.0
HBM: public
PW15iA
Connection to the mounting plate
Mounting plate
(not included in the scope of supply)
Platform center
Loading direction
Platform
(not included in the scope of supply)
Load application
(customer construction)
Connection to load application
Fig. 3.1
Mounting
Secure the load cell in the mounting holes. A platform for
load application can be mounted on top. The screws and
tightening torques to be used are given in the table
below:
Mounting
Thread
Min. property
class
Tightening
torque
Max. thread reach
On mounting plate
M6
10.9
10N
@
m
10mm
On load application
M6
10.9
10N
@
m
10mm
Important
The unevenness of the surface at the connecting faces
must be no worse than 0.1mm.