Manual parameterization and configuration
48
A00760_09_E00_00
HBM: public
ML77B
11.1.3 Parameterisation entry for 8-channel device
Byte no.
Bit no.
Meaning
Default
0
Bit 0..3
Type: 8-channel
2 (const)
0
Bit 4..7
Data format
0 = Signed16 2 bytes
1 = Signed32 4 bytes
2 = Float (IEEE format) 4 bytes
3 = Float (Siemens format) 4 bytes
4 = Raw data (0..7680000) 4 bytes
5 = Raw data (0..30000) 2 bytes
0
1
Bit0..5
Reserved
1
Bit6
Enable MOMPEAK1 control bit in
control word
0 (disabled)
1
Bit7
Enable MOMPEAK2 control bit in
control word
0 (disabled)
2
Bit 0
Enable ACAL control bit in control
word
0 (disabled)
2
Bit 1
Enable TAR control bit in control
word
0 (disabled)
2
Bit 2
Enable NULL control bit in control
word
0 (disabled)
2
Bit 3
Enable CLRPEAK1 control bit in
control word
0 (disabled)
2
Bit 4
Enable CLRPEAK2 control bit in
control word
0 (disabled)
2
Bit 5
Enable HOLDPEAK1 control bit in
control word
0 (disabled)
2
Bit 6
Enable HOLDPEAK2 control bit in
control word
0 (disabled)
3
Bit0..3
Decimal places for subchannel 1
3
Bit4..7
Decimal places for subchannel 2
4
Bit0..3
Decimal places for subchannel 3
4
Bit4..7
Decimal places for subchannel 4
5
Bit0..3
Decimal places for subchannel 5
5
Bit4..7
Decimal places for subchannel 6