![HBK CB1010 Operating Manual Download Page 46](http://html1.mh-extra.com/html/hbk/cb1010/cb1010_operating-manual_2111690046.webp)
CANHEAD
CONFIGURING WITH MGCPLUS
44
-
The strain gage measurement option must be activated by a check mark in the rele
vant column
-
Choice of compensation method (off, reference strain gage, Pt100)
-
With the reference strain gage method, its parameters must be set via the
”Enhanced settings” menu of the particular CANHEAD base module (see ”CANHEAD
configuration” on page 43)
-
With the Pt100 method, the polynomial coefficients from the strain gage packet
must be transferred. If the alpha value of the material for which the strain gage is
adjusted (
α
SG
), differs from that of the material actually available (
α
Mat
), these two
values must also be entered