![HBK CB1010 Operating Manual Download Page 36](http://html1.mh-extra.com/html/hbk/cb1010/cb1010_operating-manual_2111690036.webp)
CANHEAD
CONNECTION
34
PIN
Wire
color
Pin assignment
1
Screen
2
red
Supply Voltage (+)
3
black
Supply Voltage (-)
4
white
CAN high
5
blue
CAN low
AP74 or from another CANHEAD base module
to the next base module
LED Green
LED Red
CB1014‐350
1
2
STATUS
CE
incoming
outgoing
Fig. 7.7
Connection elements on the back
Meaning of the LEDs
The LEDs indicate the measuring point status whenever there is a ”change of status”.
When an SG is connected to measuring point 1, the green LED flashes (Status 1) once
(corresponding to the measuring point) and then stays on.
S
Green LED: OK, measuring point recognized
S
Red LED:
-
If there is an error when powering up the system with SGs connected, the red LED
flashes the number of times to correspond to the corres‐ ponding error number
(Error code, see Table below) and then stays on (flashes up to error number 32 at
maximum, i.e. 32 times).
-
The red LED also flashes in measurement mode when there is a change in status
(SG overloaded or SG connected/disconnected), to correspond to the correspond
2
3
4
1
5
1
4
3
2
5
Male connector
Female connector
Summary of Contents for CB1010
Page 1: ...CANHEAD ENGLISH DEUTSCH Operating Manual Bedienungsanleitung...
Page 3: ...CANHEAD ENGLISH DEUTSCH Operating Manual...
Page 40: ...CANHEAD CONFIGURING WITH MGCPLUS 38 S Serial number and bridge type of the connected module...
Page 52: ...CANHEAD CANHEAD ENHANCEMENTS 50...
Page 53: ...CANHEAD ENGLISH DEUTSCH Bedienungsanleitung...
Page 91: ...39 CANHEAD KONFIGURIEREN MIT MGCPLUS S Seriennummer und Br ckentyp des angeschlossenen Moduls...
Page 103: ...51 CANHEAD CANHEAD ERWEITERUNGEN...
Page 104: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A01185 13 X00 00 7 0102 0010...