36
M
AINTENANCE
AND
S
ERVICING
Warning
Disconnect power cord before
replacing a fuse.
Air Filter
Replace the air filter once a year. Mark the replacement
date on the air filter for reference.
To replace the air filter, disconnect tubing (from Purge
Pump), install replacement air filter in the correct direction,
and reconnect tubing (see Figure 21).
Check Valve
If the unit does not maintain pressure during operation, the
check valve may not be seated properly. To reseat the
check valve:
1.
Run the unit in Purge for one minute to loosen
the check valve ball.
2.
Run a test cycle. It should reseat the check valve
ball.
3.
If the check valve continues to leak under
pressure, replace it (see Figure 21).
Electrical Components
Fuses
If the control panel does not activate when Power Switch is
ON, check the fuse cartridges. If a fuse is blown, replace it
with another of the same rating. (See Wiring Diagrams.)
Model
Fuse
Code Part No.
EC5500
100V, 12 A, Time Lag T
5120-0036
115V, 10 A, Time Lag
T
261575
230V, 7 A, Time Lag
T
440-0020
EC6000
115V, 15 A, Time Lag
T
264305
110V, 15 A, Time Lag
T
264305
230V, 8 A, Time Lag
T
FZX67
Summary of Contents for Chemiclave EC5500
Page 15: ...15 Figure 4 Metering Valve Function PRINCIPLES OF OPERATION ...
Page 16: ...16 Figure 7 Pressurization PRINCIPLES OF OPERATION Figure 6 Ready Load Figure 5 Warm Up ...
Page 17: ...17 Figure 8 EC5500 EC6000 Operation PRINCIPLES OF OPERATION ...
Page 18: ...18 Figure 11 End Unload Figure 10 Purge Figure 9 Exposure PRINCIPLES OF OPERATION ...
Page 20: ...20 Figure 13 Piping Schematic PRINCIPLES OF OPERATION ...
Page 21: ...21 Figure 14 Cycle Phase Diagram PRINCIPLES OF OPERATION ...
Page 22: ...22 Door Gasket Figure 15 Exploded View Top Cover and Door PRINCIPLES OF OPERATION ...
Page 32: ...32 MAINTENANCE AND SERVICING Figure 17 Metering Valve need new drawing ...
Page 39: ...39 MAINTENANCE AND SERVICING Figure 22 Power Supply Printed Circuit Board ...
Page 56: ...56 ...