10
G
ENERAL
D
ESCRIPTION
Exhaust
Vapors from the chamber that do not liquefy in the con-
denser proceed to the exhaust connection on the filter
housing. They exit through the Chemi-Filter into the atmo-
sphere.
Chemi-Filter
The Chemi-Filter traps the formaldehyde from the exhaust
vapors before releasing them to the atmosphere. The
Chemi-Filter should be installed before the sterilizer is
operated. It must be replaced regularly.
Pump
Propels air into the chamber during the purge phase, which
forces vapors out of the chamber through the vent port.
Air Filter
Removes contaminants from intake air before the pump
propels it into the chamber.
Check Valve
Prevents chamber back pressure from contacting the pump
and the air filter.
Electrical Components
Printed Circuit Boards (PCBs)
Power Supply Board
Transforms line power into working voltages. Acts as “input/
output” board for heater and pump.
Summary of Contents for Chemiclave EC5500
Page 15: ...15 Figure 4 Metering Valve Function PRINCIPLES OF OPERATION ...
Page 16: ...16 Figure 7 Pressurization PRINCIPLES OF OPERATION Figure 6 Ready Load Figure 5 Warm Up ...
Page 17: ...17 Figure 8 EC5500 EC6000 Operation PRINCIPLES OF OPERATION ...
Page 18: ...18 Figure 11 End Unload Figure 10 Purge Figure 9 Exposure PRINCIPLES OF OPERATION ...
Page 20: ...20 Figure 13 Piping Schematic PRINCIPLES OF OPERATION ...
Page 21: ...21 Figure 14 Cycle Phase Diagram PRINCIPLES OF OPERATION ...
Page 22: ...22 Door Gasket Figure 15 Exploded View Top Cover and Door PRINCIPLES OF OPERATION ...
Page 32: ...32 MAINTENANCE AND SERVICING Figure 17 Metering Valve need new drawing ...
Page 39: ...39 MAINTENANCE AND SERVICING Figure 22 Power Supply Printed Circuit Board ...
Page 56: ...56 ...