34
M
AINTENANCE
AND
S
ERVICING
Warning
Burn Hazard: Allow the unit to cool to
room temperature before performing
this repair; burns could result.
Figure 19: Removing the Condenser Coil
Figure 20: Door Assembly
Condenser
In models EC5500 and EC6000 the condenser has a
heatsink and a coil.
Removing the Coil
1.
Remove the cover from the unit.
2.
Disconnect the fitting at the heatsink (see
Figure 19). Disconnect the tubing from the
other end of the coil.
3.
Remove the hardware that secures the coil to
the rear of the cabinet.
Condensate Tank
If the condensate tank leaks, replace it.
Door
Door Adjustment
See “Door Adjustment.”
Gasket Replacement
The door gasket seals the chamber at sterilization
pressure to a minimum of 40 psig (276 kPa).
To replace the door gasket, see “Gasket Replacement.”
Summary of Contents for Chemiclave EC5500
Page 15: ...15 Figure 4 Metering Valve Function PRINCIPLES OF OPERATION ...
Page 16: ...16 Figure 7 Pressurization PRINCIPLES OF OPERATION Figure 6 Ready Load Figure 5 Warm Up ...
Page 17: ...17 Figure 8 EC5500 EC6000 Operation PRINCIPLES OF OPERATION ...
Page 18: ...18 Figure 11 End Unload Figure 10 Purge Figure 9 Exposure PRINCIPLES OF OPERATION ...
Page 20: ...20 Figure 13 Piping Schematic PRINCIPLES OF OPERATION ...
Page 21: ...21 Figure 14 Cycle Phase Diagram PRINCIPLES OF OPERATION ...
Page 22: ...22 Door Gasket Figure 15 Exploded View Top Cover and Door PRINCIPLES OF OPERATION ...
Page 32: ...32 MAINTENANCE AND SERVICING Figure 17 Metering Valve need new drawing ...
Page 39: ...39 MAINTENANCE AND SERVICING Figure 22 Power Supply Printed Circuit Board ...
Page 56: ...56 ...