![HART Proline Promag 55 Manual Download Page 55](http://html1.mh-extra.com/html/hart/proline-promag-55/proline-promag-55_manual_601405055.webp)
Device Functions Proline Promag 55
4 Block QUICK SETUP
Hauser
55
6.2
Group HANDLING TOTALIZER
TOTALIZER
D
⇒
TOTALIZER 1
DAA
⇓
TOTALIZER 2
DAB
⇓
TOTALIZER 3
DAC
⇓
HANDLING TOTALIZER
DJA
⇒
Handling totalizer functions
Function description
TOTALIZER
➞
HANDLING TOTALIZER
➞
Handling totalizer functions
RESET ALL TOTALIZERS
(3800)
Use this function to reset the totals (including all overflows) of the totalizers (1 to 3) to
zero (= RESET).
Options:
NO
YES
Factory setting:
NO
!
Note!
If the device is equipped with a status input and if it is appropriately configured, a reset
for the totalizer (1 to 3) can also be triggered by a pulse (see function ASSIGN STATUS
INPUT (5000) on Page 105).
FAILSAFE MODE
(3801)
Use this function to define the common response of all totalizers (1 to 3) in case of error.
Options:
STOP
The totalizer is paused until the fault is rectified.
ACTUAL VALUE
The totalizer continues to count is based on the actual flow measuring value.
The fault is ignored.
HOLD VALUE
The totalizer continues to count the flow that is based on the last valid flow measuring
value (before the fault occurred).
Factory setting:
STOP
Summary of Contents for Proline Promag 55
Page 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Page 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Page 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Page 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...