![HART Proline Promag 55 Manual Download Page 138](http://html1.mh-extra.com/html/hart/proline-promag-55/proline-promag-55_manual_601405138.webp)
10 Block SPECIAL FUNCTION
Device Functions Proline Promag 55
138
Hauser
10.1.8
Function group VOLUME FLOW
SPECIAL FUNCTION
H
⇒
ADV. DIAGNOSTICS
HEA
⇒
CONFIGURATION
750
⇓
ACQUISITION
751
⇓
CONFIG. COATING
752
⇓
COATING E1
753
⇓
COATING E2
754
⇓
ELECTR. POTENTIAL 1
755
⇓
ELECTR. POTENTIAL 2
756
⇓
VOLUME FLOW
757
Function description
SPECIAL FUNCTION
➞
ADV. DIAGNOSTICS
➞
VOLUME FLOW
What is meant here by "volume flow" is the volume flow value that was acquired immediately before the test pulses were
applied to the measuring electrodes. This value serves as another basis for the interpretation of decay time constants or
electrode potentials with regard to coating formation, abrasion or corrosion.
REFERENCE VALUE
(7570)
Use this function to view the reference value for the volume flow.
User interface:
5-digit floating-point number, including unit
ACTUAL VALUE
(7571)
Use this function to view the actual measured volume flow.
User interface:
5-digit floating-point number, including unit
MINIMUM VALUE
(7572)
Use this function to view the lowest measured value for the volume flow, since the last
reset or deletion of the stored values.
User interface:
5-digit floating-point number, including unit
MAXIMUM VALUE
(7573)
Use this function to view the highest measured value for the volume flow, since the last
reset or deletion of the stored values.
User interface:
5-digit floating-point number, including unit
HISTORY
(7574)
Use this function to view the last 10 measured values for the volume flow.
User interface:
5-digit floating-point number, including unit
ACTUAL DEVIATION
(7575)
Use this function to view the deviation between the actual (last measured) value for the
volume flow and the reference values selected in the SELECT REFERENCE CONDITION
function (7502).
User interface:
5-digit floating-point number, including unit
Summary of Contents for Proline Promag 55
Page 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Page 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Page 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Page 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...