![Hardy LC300 Installation And Operating Instructions Manual Download Page 51](http://html1.mh-extra.com/html/hardy/lc300/lc300_installation-and-operating-instructions-manual_3911318051.webp)
HARDY MANUFACTURING CO., INC (MODEL LC300, LC500 & LC2500)
PAGE 37
8-1
GENERAL TROUBLE SHOOTING GUIDE
Complaint
Area of Trouble
Possible Causes
Corrective Action
No Heat
Main power
supply and
ground fault
interrupter
receptacle
“OK”
Water
temperature
is “HOT”
Pump performance
Unplug pump and bypass relay by tying relay
wires #4 and #6 together. Plug pump back in.
Pump does not run
1. Defective pump
cartridge
2. Defective pump
capacitor
3. Defective pump
motor winding
1. Unplug pump. Remove
pump. Pull cartridge
from pump and spin
impeller by hand. If
impeller will not spin,
replace cartridge.
2. Use ohm meter. When
the meter is connected
to the capacitor, the
needle should jump
towards “0” ohms and
slowly drift back to
infinity. Replace if
defective.
3. Disconnect the wires
connected to the motor
terminals to test the
motor independent of
electrical connections.
Check for ground and
continuity with ohm
meter. The insulator of
the windings should
show no breakage. If it
does, replace pump