Hardy Floor Scales
25
Step 15. Run the load sensor cable through the
channel to the summing box. Bundle up
any excess cable and store it in the chan-
nel. DO NOT CUT THE CABLE TO
MAKE IT FIT BETTER.
If you cut the
cable, you will not get correct readings
from the load sensors and it will void
your warranty.
Step 16. Reconnect the cable wires according to the
color code in Step 3 to the correct load
sensor connector.
WARNING
LOAD CELL CABLE LENGTH HAS BEEN
CALCULATED INTO C2 CALIBRATION DATA.
HARDY RECOMMENDS THAT YOU DO NOT
CUT YOUR ADVANTAGE LOAD SENSOR
CABLE, AS YOUR C2 ACCURACY WILL BE
AFFECTED AND THE WARRANTY WILL BE
VOIDED.
Step 17. Reassemble the Summing Box Chassis by
replacing the top cover and installing and
tightening the 8 hex head screws using the
5.5mm hex wrench or a 5.5mm hex socket.
Tighten to 10 in-lbs.
OPTIONAL PIT
FRAMES
About Pit Frames
The pit frame optional accessory is a one-piece
welded unit with no additional welding required. This
accessory is designed for in-floor or 'flush' applica-
tions. In general, a hole is cut in the concrete, the pit-
frame accessory is installed in the hole, then concrete
is poured around and under the frame. Once cured, the
scale platform is set into the frame and installation
can be completed.
Standard duty frames are available in carbon steel or
stainless steel for all the floor scale sizes. The con-
crete work and frame setting is usually done by a con-
tractor, with a scale technician completing the project
by setting and installing the scale.