Hardy Floor Scales
12
Step 3.
Connect the C2 Certified Summing Card
Interface Cable to the weight instrument.
Color code wires as follows:
•
+ Excitation
Red
•
+ Sense
Blue
•
+ Signal
Green
•
- Signal
White
•
- Sense
Brown
•
- Excitation
Black
•
+ C2
Grey
•
- C2
Violet
Step 4.
Use the weight instrument/indicator man-
ual for calibration instructions.
NOTE:
To purchase C2
®
certified cable, contact your local
Hardy Representative or the Hardy Process Solutions
Service Center.
C2
®
Second
Generation
Calibration
Definition - C2
®
electronically calibrates a scale sys-
tem without the need for test weights. This feature is
built into all Hardy weight or rate controllers, and in
the Hardy weight modules for Rockwell PLCs.
Step 1.
Refer to the Hardy weight or rate control-
ler or the Rockwell PLC module manual
for complete calibration instructions.
Test Weight
Calibration
(Hard Cal)
Requirements:
•
Test weight calibration requires the use of
certified class ‘F’ test weights equal to a
minimum of 80% of the rated scale capacity.
•
Three weights between 10% and 100% of
the scale capacity should be used to check
the mid range.
•
Several low capacity weights equivalent to
one or two instrument divisions are required
to check the systems' sensitivity.