Chapter 5 – Setting Scale Parameters
Hardy Process Solutions HI 6200 Weigh Processor User Guide
59
5.2.11 WAVESAVER®
There are six selectable levels. 0 provides no vibration immunity with the fastest response
time, while 5 provides the most vibration immunity with the slowest response time. Default
setting is 3.
Table 5-2. WAVESAVER® Levels
Immunity
Setting
Off
0
7.5 Hz
1
3.5 Hz
2
1.0 Hz
3 (default)
0.50 Hz
4
0.25 Hz
5
WARNING: BINDING ON A SCALE/VESSEL OR LOAD CELL CAN DENY THE
LOAD CELL FREE VERTICAL MOVEMENT AND PREVENT THE INSTRUMENT
FROM RETURNING TO THE ORIGINAL ZERO REFERENCE POINT.
ATTENTION: LIEU SUR UNE ÉCHELLE / UN BATEAU OU UNE CELLULE DE
CHARGE PEUT ENDOMMAGER LE MOUVEMENT VERTICAL LIBRE DE LA
CHARGE DE CHARGE ET ÉVITER L'INSTRUMENT DE RETOURNER AU
POINT DE RÉFÉRENCE ZERO ORIGINAL.