Chapter 4 –
Ways to Set
Scale Parameters
Hardy Process Solutions HI 6200 Weigh Processor User Guide
50
13. Click
OK
to accept the value, and then click the
Send
button to send the values to the HI
6200.
14. When the Response OK message appears, click
OK
.
15. Change the display back to Float and confirm that the Tare command changed the Net
weight to zero.
4.3.6.2 I/O Tables for Communications to PLC
Table 4-6 and Table 4-7 on page 51 are common for the EtherNet/IP and Modbus – TCP
communication protocols.
Table 4-6. Output Table
Output Table
Type
Command
INT
Aux Command Information
INT
Parameter Value
DINT
Parameter ID
INT
Reserved 1
INT
Reserved 2
INT
Reserved 3
INT
Reserved 4
INT
Reserved 5
INT
Total
10
Command
,
Aux Command Information
,
Parameter ID
, and
Parameter Value
are used
to send commands to the instrument, write new parameter values, read existing parameter