Hardy Process Solutions HI 6200 Weigh Processor User Guide
52
A zero indicating the command passed.
A value of 0xFF indicating the command is in process.
An error code indicating the reason the command failed (see Table 4-5).
The upper 8 bits of the register are a cyclic “measurement update count,” which increment
by 1 (one) each time a new measurement value is taken, following a 0 to 255 then repeat
cycle. If this value remains the same in two consecutive reads from the instrument, the
communication or the measurement function has failed and appropriate action must be
taken.
Parameter ID
is an echo of the value sent in the output table.
Parameter Value
is the
current value for the specified Parameter ID. This parameter value may ether be an integer
or in floating point format depending on which Parameter ID is specified.
Instrument Status
is a 16-bit value that provides the current state of all the major
functions within the instrument. They reflect the status of all the major functions and should
be used with the “measurement update count” to determine the health of the instrument.
The
Net Weight
and
Gross
Weight
values are always provided.
4.3.6.3 Default Parameter Values
Table 4-8 shows the default parameters values.
Table 4-8. Default Parameters Values
Configuration Table
Data Type
Default
Range
Values
Gravity Correction
FLOAT
1.0
0.9-1.1
Decimal Point
BYTE
2
0-5
Grad Size
BYTE
0
0-9
0 = 1
1 = 2
2 = 5
3 = 10
4 = 20
5 = 50
6 = 100
7 = 200
8 = 500
9 = 1000
Motion Tolerance
FLOAT
10.0
.00001-999999.99
Zero Tolerance
FLOAT
10.0
.00001-999999.99
Tare Weight
FLOAT
0.0
Reference Weight
FLOAT
0.0
.00001-999999.99
Scale Cap
FLOAT
999999.0
.00001-999999.99
Span Weight
FLOAT
1000.0
.00001-999999.99