F. Verify Individual Load Sensor Millivolt Output readings
Testing an individual load sensor output requires an IT summing card or Millivolt meter with two decimal place
resolution.To determine the sensitivity and parameters for your load sensor, use the load sensor certificate or read the C2
chip with the utility found under the Diagnostic menu.
Example: A 3mV/V Load sensor produces approximately 3.0000 mV/V at rated load. If the scale capacity is 1,000 lbs
with 100 lbs of deadload, when empty the load point mV/V reading should measure 0.3000 mV/V. This number can be
converted to mV by multiplying the mV/V results by 5. (5V excitation)