HI 6600 Series Modular Sensor System User Guide
Page | 13
DC Input Voltage Hardy Gateway Module (HGM)
3.90 Watts, 12-24 VDC, 162-290 mA
Optional HI 6110 Front Panel Display
Monochrome 480 x 272 LCD display with back light
Five tactile keys for menu item selection
Displays in either white on black or black on white
IP66 rated when mounted on a smooth, rigid surface (minimum 16-gage) using the supplied gasket
2 watts maximum
Environmental Requirements
Operating Temperature Range
-10ºC to 60º C (14º to 140º F)
Temperature Coefficient
Less than 0.005% of full scale per degree C for Cal-LO and Cal-HI reference points
Storage Temperature Range
-40 to 85º C (-40º to 185º F)
Humidity Range
0-90% (non-condensing)
Environmental
Intended for Building-in, indoor use only at ambient temperatures between 10ºC to 60º C (14º to 140º
F) with a pollution degree of 2.
Approvals
UL, CUL and CE
Network Certifications
EtherNet/IP ODVA Conformance Tested, Level 3
Profibus-DP by Profibus.org
Features and Capabilities
Hardy Process Toolbox
The Hardy Process Toolbox is a set of productivity tools that support process weighing functions. Each tool in
the Hardy Process Toolbox saves time, increases accuracy, improves efficiency or reduces risk in process
weighing applications. The HI 6600 includes the Toolbox functions discussed below.
C2
®
Calibration
Traditional calibration uses certified test weights. C2® Electronic Calibration allows a scale to be calibrated
without the need for test weights. A C2 weighing system consists of up to four load cell sensors per WPM, a
junction box, interconnect cable, and an instrument with C2 capabilities (e.g., the HI 6600 series instrument).
Each Hardy Process Solutions C2-certified load sensor outputs digital information used for calculating the
Summary of Contents for HI 6110
Page 16: ...HI 6600 Series Modular Sensor System User Guide Page 16 ...
Page 44: ...HI 6600 Series Modular Sensor System User Guide Page 44 ...
Page 97: ...HI 6600 Series Modular Sensor System User Guide Page 97 A ElectricalandMechanicalReview ...
Page 107: ...HI 6600 Series Modular Sensor System User Guide Page 107 H Mechanical Installation ...
Page 108: ...HI 6600 Series Modular Sensor System User Guide Page 108 J Electrical Inspection ...
Page 109: ...HI 6600 Series Modular Sensor System User Guide Page 109 K InstallationCheckPoints ...
Page 128: ...HI 6600 Series Modular Sensor System User Guide Page 128 ...