[Fig 57]
Common waiting operation
Name
PTN NO.
Temp. WAIT
RANGE
Humi. WAIT
RANGE
WAIT TIME
Function
Enter a pattern number to be set or select it by
pressing the
/ button.
Set the deviation range about the temperature set
value which will be applied to the standby operation.
Set the deviation range about the humidity set value
which will be applied to the standby operation.
Until it satisfies the deviation range about the set
value which will be applied to the standby operation,
it sets the time that delays the process of segment.
When both of temperature and humidity standby
range are being set, both of them must fall within
standby operation range in order to cancel the
standby operation.
Range
1~100 Pattern
(0 ~ ±300) ℃(TH500)
(0 ~ ±600) ℃(TH300)
(0 ~ ±100) %
0 ~ 99 hour
59 minute
Waiting time
Waiting Range
Standby operation
becomes cancelled because PV
falls within the standby range
S
b
t
s
d
b
Standby operation starts
to on because PV did not fall
within the standby range
S
w
Start Waiting
Operation
S
SEG progress without
applying standby mode
SEG progress with
applying standby mode
S
a
a
50
[FIG 57] generally display the standby operation. If process value (PV) does not fall within the
standby range at the changing point from SEG1 to SEG2, it will wait until process value (PV)
falls within the standby range (wait for the set standby time). Here, if standby time elapses, it
will process to the SEG 2 even if the process value (PV) did not fall within the standby range
[FIG 58].
th300_500_통합본_th300_500_통합본.qxp 19. 6. 26. 오후 1:34 페이지 50