[FIG.118] D.O Configuration Setting 7
It shows an example of displaying either Relay or O/C output corresponding to the temperature
UP/HOLD/DOWN setting of temperature and humidity SV when operating Program Control. Ex :
section of temp.UP(-10 ℃), section of temp.HOLD(2 Min) and section of temp DOWN(+20℃) It
indicates the ON timing under the assigned Relay or O/C output.
Section of
temp. Up
Up Setting
Section of
temp. Down
Section of
temp. Hold
Section of
temp. Hold(3 minute)
Hold setting
=2minute
Down setting
=20
¡
Relay or
O/C Output
Relay or
O/C Output
Relay or
O/C Output
[Fig.119] Output by temperature UP/HOLD/DOWN setting
D.O Configuration setting 7 [FIG 118] is used when turning ON the output within the
increase/maintain/decrease interval of each of temperature/humidity set value (SV). Each set
input value of setting lists operates same as [FIG 119]
Explanation
Input the minus value of Temp. & Humi. in the target of setting value
Input the plus value of Temp. & Humi. in the target of setting value
Input the holding time of Relay or O.C output within segment time
Setting item
Temp. & Humi. Up
Temp. & Humi. Down
Temp. & Humi. SOAK
6.7.6 D.O Configuration Setting 7