4.
Press the Right Key at the Calibrate Output 1A prompt. At the 20.000mA prompt
press Right Key once. Use the Up Key or the Down Key to adjust the display to
the reading on the volt/ammeter. The controller should stabilize within one
second. Repeat until the volt/ammeter reads 20.000mA. Press Right Key to
store the value.
Output 1B Milliamperes (Not Required)
Equipment
•
Precision volt/ammeter with 3.5-digit resolution.
Setup and Calibration
1.
Connect the correct power supply to terminals 1, 2 and 3 (see the Wiring
Chapter and the Appendix).
2.
Connect the volt/ammeter to terminals 39 (+) and 40 (-).
3.
Press the Right Key at the Calibrate Output 1B prompt. At the 4.000mA prompt
press Right Key once. Use the Up Key or the Down Key to adjust the display to
the reading on the volt/ammeter. The controller should stabilize within one
second. Repeat until the volt/ammeter reads 4.000mA. Press Right Key to store
the value.
4.
Press the Right Key at the Calibrate Output 1B prompt. At the 20.000mA prompt
press Right Key once. Use the Up Key or the Down Key to adjust the display to
the reading on the volt/ammeter. The controller should stabilize within one
second. Repeat until the volt/ammeter reads 20.000mA. Press Right Key to
store the value.
Output 2A Milliamperes (Not Required)
Equipment
•
Precision volt/ammeter with 3.5-digit resolution.
Setup and Calibration
1.
Connect the correct power supply to terminals 1, 2 and 3 (see the Wiring
Chapter and the Appendix).
2.
Connect the volt/ammeter to terminals 36 (+) and 37 (-).
3.
Press the Right Key at the Calibrate Output 2A prompt. At the 4.000mA prompt
press Right Key once. Use the Up Key or the Down Key to adjust the display to
the reading on the volt/ammeter. The controller should stabilize within one
second. Repeat until the volt/ammeter reads 4.000mA. Press Right Key to store
the value.
4.
Press the Right Key at the Calibrate Output 2A prompt. At the 20.000mA prompt
press Right Key once. Use the Up Key or the Down Key to adjust the display to
the reading on the volt/ammeter. The controller should stabilize within one
second. Repeat until the volt/ammeter reads 20.000mA. Press Right Key to
store the value.
Authorization and Copyright of this document is
controlled by Hanse Environmental Inc.
25
Summary of Contents for VTC Series
Page 1: ...Operation and Maintenance Instructions for Hanse Environmental Inc VTC Series Rev H 2018 ...
Page 42: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 42 ...
Page 56: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 56 ...
Page 57: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 57 ...
Page 58: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 58 ...
Page 59: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 59 ...
Page 60: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 60 ...
Page 61: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 61 ...
Page 64: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 64 ...
Page 65: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 65 ...
Page 66: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 66 ...
Page 69: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 69 ...
Page 70: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 70 ...
Page 71: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 71 ...
Page 79: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 79 ...
Page 83: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 83 ...