METHOD PROCEDURES

42 

42

CHLORINE

, FREE

10.5.  CHLORINE, FREE

SPECIFICATIONS

Range 

0.00 to 5.00 mg/L (as Cl

2

)

Resolution 

0.01 mg/L

Accuracy 

±0.03 mg/L ±3% of reading at 25 °C

Light Source 

LED with narrow band interference filter @ 525 nm

Method 

Adaptation of the EPA DPD Method 330.5 

REQUIRED REAGENTS

POWDER

Code Description 

Quantity

HI93701-0 

Free Chlorine Reagent 

1 packet

LIQUID

Code Description 

Quantity

HI93701A-F

 

Free Chlorine Reagent A 

3 drops

HI93701B-F

 

Free Chlorine Reagent B 

3 drops

REAGENT SETS

HI93701-F 

Reagents for 300 tests (liquid)

HI93701-01

 

Reagents for 100 tests (powder)

HI93701-03

 

Reagents for 300 tests (powder)

For other accessories see ACCESSORIES section.

MEASUREMENT PROCEDURE

• Select the 

Chlorine (Free) 

method using the procedure described in the METHOD SELECTION section.

• Fill the cuvette with 10 mL of unreacted sample (up to the mark). 

Replace the plastic stopper and the cap.

• Insert the cuvette into the holder and close the lid.

10 mL

Summary of Contents for HI83326

Page 1: ...INSTRUCTION MANUAL HI83326 Multiparameter Photometer for Pool Spa...

Page 2: ...nna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as wel...

Page 3: ...6 3 GENERAL SETUP 13 6 4 CONTEXTUAL HELP 16 7 LOGGING DATA DATA MANAGEMENT 17 7 1 LOGGING DATA 17 7 2 ADDING SAMPLE USER NAMES TO LOG DATA 17 7 3 DATA MANAGEMENT 18 8 PHOTOMETER MODE 19 8 1 METHOD SEL...

Page 4: ...45 10 7 COPPER HIGH RANGE 49 10 8 CYANURIC ACID 51 10 9 HARDNESS CALCIUM 53 10 10 IRON HIGH RANGE 56 10 11 NITRATE 58 10 12 OZONE 61 10 13 pH 65 10 14 PHOSPHATE LOW RANGE 67 11 WARNINGS ERRORS 69 12 S...

Page 5: ...re sure that the instrument works correctly Any damaged or defective item must be returned in its original packing material with the supplied accessories 2 SAFETY MEASURES The chemicals contained in t...

Page 6: ...vailable buffers 4 01 6 86 7 01 9 18 10 01 pH Electrode Intelligent pH temperature electrode Temperature Range 20 0 to 120 0 C 4 0 to 248 0 F Resolution 0 1 C 0 1 F Accuracy 0 5 C 25 C 0 9 F 77 F Addi...

Page 7: ...e GLP Good Laboratory Practice NIST National Institute of Standards and Technology EPA US Environmental Protection Agency g L grams per liter ppt g L micrograms per liter ppb mg L milligrams per liter...

Page 8: ...ceable standards Probe mode uses a digital pH probe with a one or two point calibration Digital electrode input for pH measurements Certified CAL Check cuvettes to confirm meter functionality Dual pur...

Page 9: ...LCD 3 Indexing mark 4 Protective port covers 5 Light blocking cover panel 6 Cuvette holder 7 ON OFF power button 8 3 5 mm TRRS jack input for digital electrodes 9 Standard USB host connector for data...

Page 10: ...up screen Press to log the current reading Press to review saved logs Press to exit the current screen Press to display the help screen ON OFF power button 5 4 PRINCIPLE OF OPERATION Absorption of lig...

Page 11: ...ment and sample measurement LED light sources offer superior performance compared to tungsten lamps LEDs have a much higher luminous efficiency providing more light while using less power They also pr...

Page 12: ...to AC DC adapter battery capacity no external adapter battery near 0 no external adapter battery exhausted no external adapter To conserve battery the meter will turn off automatically after 15 minute...

Page 13: ...al information Temperature Unit pH Only Option C or F Press the functional key to select the desired temperature unit Backlight Values 0 to 8 Press Modify to access the backlight intensity Use the fun...

Page 14: ...4 hour Press the functional key to select the desired time format Date Format Option DD MM YYYY MM DD YYYY YYYY MM DD YYYY MM DD Mon DD YYYY DD Mon YYYY YYYY Mon DD Press Modify to change the date for...

Page 15: ...Press the functional key to enable or disable the beeper Instrument ID Option 0 to 999999 This option is used to set the instrument s ID identification number Press Modify to access the instrument ID...

Page 16: ...rn to the Setup menu 6 4 CONTEXTUAL HELP HI83326 offers an interactive contextual help mode that assists the user at any time To access the help screen press the HELP key The instrument will display a...

Page 17: ...ded to the saved log Use the keys to highlight the Sample ID or User ID then press Modify Sample ID and user ID are entered using the alphanumeric multi tapping keypad Enter one character at a time by...

Page 18: ...l Data Export Log data can be exported to a USB flash drive or to a PC To access data export functions press the RECALL key then Export Use the keys to select the desired export location For export to...

Page 19: ...thod read all the instructions carefully 8 2 COLLECTING MEASURING SAMPLES AND REAGENTS Proper Use of Syringe 1 Push the plunger completely into the syringe and insert the tip into the solution 2 Pull...

Page 20: ...ortant for reproducibility of the measurements The proper mixing technique for each method is listed in the method procedure a Invert the cuvette a couple of times or for a specified time hold the cuv...

Page 21: ...ack cap Whenever the cuvette is placed into the measurement holder it must be dry outside and free of fingerprints oil and dirt Wipe it thoroughly with HI731318 microfiber cleaning cloth or a lint fre...

Page 22: ...e than one timer the meter will automatically select each timer in the appropriate order To bypass the default order you may press the desired key to activate a different timer only while the current...

Page 23: ...of each CAL Check Standard and applies the factory calibration corrections to each measurement The HI83326 stores the results of the most recent CAL Check measurements which may be viewed on the CAL...

Page 24: ...D key 2 Highlight the appropriate Absorbance method according to the wavelength to be used then press Select Absorbance methods are located at the bottom of the method list 3 Prepare the sample cuvett...

Page 25: ...tion for more information To take pH measurements Remove the protective cap and rinse the electrode with water Collect some sample in a clean dry beaker Preferably rinse the electrode with a small amo...

Page 26: ...nd pH 4 01 as the second buffer If you are measuring in the alkaline range use pH 7 01 or 6 86 as the first buffer and pH 10 01 or 9 18 as the second buffer Procedure Calibration can be performed usin...

Page 27: ...be Exceeded Probe Range ThepHand ortemperaturemeasurementexceedthespecifications of the probe The measurement value s will be blinking Broken Temperature Sensor The temperature sensor inside the probe...

Page 28: ...ptable window for the selected buffer value 9 4 pH GLP Good Laboratory Practice GLP refers to a quality control function used to ensure uniformity and consistency of sensor calibrations and measuremen...

Page 29: ...electrodes if the filling solution electrolyte is more than 2 cm 1 below the fill hole addHI7082orHI80823 5MKClElectrolytesolution Unscrewthefillholecoverduringmeasurements so the liquid reference jun...

Page 30: ...olution for approximately 30 minutes Protein Soak in Hanna HI7073 or HI8073 Protein cleaning solution for 15 minutes Inorganic Soak in Hanna HI7074 Inorganic cleaning solution for 15 minutes Oil and g...

Page 31: ...alinity Reagent 1 mL HI93755 53 Chlorine Removal Reagent 1 drop REAGENT SETS HI775 26 Reagents for 25 tests For other accessories see ACCESSORIES section MEASUREMENT PROCEDURE Select the Alkalinity me...

Page 32: ...and the cap Invert 5 times Insert the cuvette into the holder and close the lid Press Read to start the reading The instrument displays the results in mg L calcium carbonate CaCO3 INTERFERENCES Inter...

Page 33: ...e Description Quantity HI93716 0 Bromine Reagent 1 packet REAGENT SETS HI93716 01 Reagents for 100 tests HI93716 03 Reagents for 300 tests For other accessories see ACCESSORIES section MEASUREMENT PRO...

Page 34: ...Read When the timer ends the meter will perform the reading The instrument displays the results in mg L of bromine Br2 INTERFERENCES Interference may be caused by Chlorine Iodine Ozone Oxidized forms...

Page 35: ...L HI93738D 0 Chlorine Dioxide Reagent D 1 mL REAGENT SETS HI93738 01 Reagents for 100 tests HI93738 03 Reagents for 300 tests For other accessories see ACCESSORIES section SAMPLING PROCEDURE It is rec...

Page 36: ...al times to mix 1 0 5 mL 2 0 5 mL 1 2 Add one packet of HI93738B 0 Chlorine Dioxide Reagent B to one of the two cylinders 1 cap and invert it several times until it is totally dissolved This is the bl...

Page 37: ...mL 1 2 Fill cuvette 1 with 10 mL of the blank up to the mark Replace the plastic stopper and the cap Insert the blank 1 into the holder and close the lid Press Zero The display will show 0 0 when the...

Page 38: ...METHOD PROCEDURES 38 38 CHLORINE DIOXIDE Press Read to start the reading The instrument displays the results in mg L of chlorine dioxide ClO2 INTERFERENCES Interference may be caused by Strong oxidan...

Page 39: ...e sample the addition of glycine as a masking agent inhibits the response of free chlorine APPLICATION Drinking water tap water treated water SAMPLING PROCEDURE Collect the sample in a clean glass bot...

Page 40: ...ELECTION section Fill the cuvette with 10 mL of unreacted sample up to the mark Add 5 drops of HI96779A 0 Chlorine Dioxide Reagent A Replace the plastic stopper and the cap Shake gently for 30 seconds...

Page 41: ...w the countdown prior to the measurement or wait 1 minute and press Read When the timer ends the meter will perform the reading The instrument displays the results in mg L of ClO2 INTERFERENCES Interf...

Page 42: ...Code Description Quantity HI93701A F Free Chlorine Reagent A 3 drops HI93701B F Free Chlorine Reagent B 3 drops REAGENT SETS HI93701 F Reagents for 300 tests liquid HI93701 01 Reagents for 100 tests...

Page 43: ...one packet of HI93701 0 Free Chlorine Reagent Replace the plastic stopper and the cap Shake gently for 20 seconds Insert the cuvette into the holder and close the lid Press Timer and the display will...

Page 44: ...esults in mg L of chlorine Cl2 Note Free and Total Chlorine have to be measured separately with fresh sample following the related procedure if both values are desired INTERFERENCES Interference may b...

Page 45: ...HI93701A T Total Chlorine Reagent A 3 drops HI93701B T Total Chlorine Reagent B 3 drops HI93701C T Total Chlorine Reagent C 1 drop REAGENT SETS HI93701 T Reagents for 300 tests liquid HI93711 01 Reage...

Page 46: ...93711 0 Total Chlorine Reagent Replace the plastic stopper and the cap Shake gently for 20 seconds Insert the cuvette into the holder and close the lid Press Timer and the display will show the countd...

Page 47: ...e up to the mark Replace the plastic stopper and the cap Shake gently Insert the cuvette into the holder and close the lid Press Timer and the display will show the countdown prior to the measurement...

Page 48: ...greater than 500 mg L CaCO3 to remove the interference shake the sample for approximately 2 minutes after adding the powder reagent Alkalinity greater than 250 mg L CaCO3 or acidity greater than 150...

Page 49: ...Copper High Range Reagent 1 packet REAGENT SETS HI93702 01 Reagents for 100 tests HI93702 03 Reagents for 300 tests For other accessories see ACCESSORIES section MEASUREMENT PROCEDURE Select the Copp...

Page 50: ...the holder and close the lid Press Timer and the display will show the countdown prior to the measurement or wait 45 seconds and press Read When the timer ends the meter will perform the reading The i...

Page 51: ...Acid Reagent 1 packet REAGENT SETS HI93722 01 Reagents for 100 tests HI93722 03 Reagents for 300 tests For other accessories see ACCESSORIES section MEASUREMENT PROCEDURE Select the Cyanuric Acid met...

Page 52: ...L of the reacted sample up to the mark Replace the plastic stopper and the cap Insert the cuvette into the holder and close the lid Press Timer and the display will show the countdown prior to the mea...

Page 53: ...alcium Hardness Reagent B 0 5 mL HI93720C 0 Calcium Hardness Reagent C 1 drop REAGENT SETS HI93720 01 Reagents for 100 tests HI93720 03 Reagents for 300 tests For other accessories see ACCESSORIES sec...

Page 54: ...vette several times to mix This is the blank Insert the blank 1 into the holder and close the lid Press Zero The display will show 0 0 when the meter is zeroed and ready for measurement Remove the bla...

Page 55: ...iness repeat the test multiple times until you obtain consistent results SAMPLE DILUTION This meter is designed to determine low levels of hardness typically found in water purification systems Diluti...

Page 56: ...RED REAGENTS Code Description Quantity HI93721 0 Iron High Range Reagent 1 packet REAGENT SETS HI93721 01 Reagents for 100 tests HI93721 03 Reagents for 300 tests For other accessories see ACCESSORIES...

Page 57: ...vette into the holder and close the lid Press Timer and the display will show the countdown prior to the measurement or wait 3 minutes and press Read When the timer ends the meter will perform the rea...

Page 58: ...HI93728 0 Nitrate Reagent 1 packet REAGENT SETS HI93728 01 Reagents for 100 tests HI93728 03 Reagents for 300 tests For other accessories see ACCESSORIES section MEASUREMENT PROCEDURE Select the Nitra...

Page 59: ...er will not completely dissolve Note Themethodistechniquesensitive Seeproceduredescribed in CUVETTE PREPARATION section for proper mixing technique Insert the cuvette into the holder and close the lid...

Page 60: ...te NO3 Press the or key to return to the measurement screen INTERFERENCES Interference may be caused by Ammonia and amines as urea and primary aliphatic amines Chloride above 100 mg L Chlorine above 2...

Page 61: ...1 packet REAGENT SETS HI93757 01 Reagents for 100 tests HI93757 03 Reagents for 300 tests HI93703 52 Reagents for 100 tests Optional For other accessories see ACCESSORIES section STANDARD MEASUREMENT...

Page 62: ...show the countdown prior to the measurement or wait 2 minutes and press Read When the timer ends the meter will perform the reading The instrument displays the result in mg L ozone O3 chlorine free sa...

Page 63: ...e cap Shake gently until the powder is completely dissolved Add one packet of HI93757 0 Ozone Reagent Replace the plastic stopper and the cap Shake gently for 20 seconds Insert the cuvette into the ho...

Page 64: ...er reagent Alkalinity above 250 mg L CaCO3 will not reliably develop the full amount of color or it may rapidly fade neutralize the sample with diluted HCl If the sample is suspected to contain chlori...

Page 65: ...drops REAGENT SETS HI93710 01 Reagents for 100 tests HI93710 03 Reagents for 300 tests For other accessories see ACCESSORIES section MEASUREMENT PROCEDURE Select the pH method using the procedure desc...

Page 66: ...ette and add 5 drops of HI93710 0 pH Reagent Indicator Replace the plastic stopper and the cap and mix the solution Insert the cuvette into the holder and close the lid Press Read to start the reading...

Page 67: ...ACCESSORIES section MEASUREMENT PROCEDURE Select the Phosphate LR method using the procedure described in the METHOD SELECTION section Rinseand replace the plastic stopper and the cap Shake the cuvet...

Page 68: ...erform the reading The instrument displays concentration in mg L of phosphate PO4 3 Press the or key to access the second level functions Press Chem Frm to convert the result to mg L of phosphorus P a...

Page 69: ...repeat the measurement There is either too much light or the instrument can not adjust the light level Please check the preparation of the zero cuvette and that the sample does not contain any debris...

Page 70: ...is not recognized or it might be damaged Please insert a new USB flash drive Data log is full Please review logged data and delete unnecessary logs Date and time settings have been lost Please reset...

Page 71: ...please contact Hanna Instruments technical support Logged data is no longer accessible Some features are no longer available Restart the meter If the issue persists please contact Hanna Instruments t...

Page 72: ...O2 DPD Chlorine Free 0 00 to 5 00 mg L as Cl2 DPD Chlorine Total 0 00 to 5 00 mg L as Cl2 DPD Copper HR 0 00 to 5 00 mg L as Cu 2 Bicinchoninate Cyanuric Acid 0 to 80 mg L as CYA Turbidimetric Hardnes...

Page 73: ...52 100 ozone tests HI93710 01 100 pH tests HI93710 03 300 pH tests HI93711 01 100 chlorine total tests powder HI93711 03 300 chlorine total tests powder HI93713 01 100 phosphate LR tests HI93713 03 3...

Page 74: ...y double junction refillable pH temperature electrode with enhanced diagnostics HI12300 Plastic body double junction gel filled non refillable pH temperature electrode HI12301 Plastic body double junc...

Page 75: ...solution 500 mL HI7009L pH 9 18 buffer solution 500 mL HI7010L pH 10 01 buffer solution 500 mL HI8004L pH 4 01 buffer solution in FDA approved bottle 500 mL HI8006L pH 6 86 buffer solution in FDA appr...

Page 76: ...olution in FDA approved bottle 500 mL ELECTRODE REFILL ELECTROLYTE SOLUTIONS Code Description HI7082 3 5M KCl electrolyte 4x30 mL for double junction electrodes HI8082 3 5M KCl electrolyte in FDA appr...

Page 77: ...I740229 100 mL graduated cylinder DEMI 02 demineralizer HI75110 220E USB power adapter European plug HI75110 220U USB power adapter USA plug HI76404A electrode holder HI83326 11 CAL Check cuvette kit...

Page 78: ...is used Any variation introduced by the user to the supplied equipment may degrade the photometer s performance For yours and the meter s safety do not use or store the photometer in hazardous enviro...

Page 79: ...Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice...

Page 80: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN83326 Printed in ROMANIA...

Reviews: