33
METHOD PROCEDURES
33
33
BROMINE
10.2. BROMINE
SPECIFICATIONS
Range
0.00 to 8.00 mg/L (as Br
2
)
Resolution
0.01 mg/L
Accuracy
±0.08 mg/L ±3% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 525 nm
Method
Adaptation of Standard Methods for the Examination of Water and
Wastewater, 18
th
Edition, DPD Method
REQUIRED REAGENTS
Code Description
Quantit
y
HI93716-0
Bromine Reagent
1 packet
REAGENT SETS
HI93716-01
Reagents for 100 tests
HI93716-03
Reagents for 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the
Bromine
method using the procedure described in the METHOD SELECTION section.
• Fill the cuvette with 10 mL of unreacted sample (up to the mark).
Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and close the lid.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
10 mL