51
METHOD PROCEDURES
51
51
CY
ANURIC ACID
10.8. CYANURIC ACID
SPECIFICATIONS
Range
0 to 80 mg/L (as CYA)
Resolution
1 mg/L
Accuracy
±1 mg/L ±15% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 525 nm
Method
Adaptation of the Turbidimetric Method
REQUIRED REAGENTS
Code Description
Quantity
HI93722-0
Cyanuric Acid Reagent
1 packet
REAGENT SETS
HI93722-01
Reagents for 100 tests
HI93722-03
Reagents for 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the
Cyanuric Acid
method using the procedure described in the METHOD SELECTION section.
• Fill the first cuvette with 10 mL of unreacted sample (up to
the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and close the lid.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
10 mL
# 1
# 1