METHOD PROCEDURES
63
63
OZONE
• Press
Zero
. The display will show “-0.0-” the meter is zeroed and ready for measure.
• Remove the cuvette.
• Add one packet of the
HI93703-52-0
Glycine Powder.
Replace the plastic stopper and the cap. Shake gently until
the powder is completely dissolved.
• Add one packet of
HI93757-0
Ozone Reagent. Replace the
plastic stopper and the cap. Shake gently for 20 seconds.
• Insert the cuvette into the holder and close the lid.
• Press
Timer
and the display will show the countdown prior to the measurement or wait 2 minutes
and press
Read.
When the timer ends the meter will perform the reading. Record this value as B.