METHOD PROCEDURES
40
40
CHLORINE DIO
XIDE
, RAPID METHOD
MEASUREMENT PROCEDURE
• Select the
Chlorine Dioxide (Rapid)
method
using the procedure described in the METHOD
SELECTION section.
• Fill the cuvette with 10 mL of unreacted sample (up to the mark).
• Add 5 drops of
HI96779A-0
Chlorine Dioxide Reagent A.
• Replace the plastic stopper and the cap. Shake gently for
30 seconds.
• Wait 30 seconds.
• Insert the cuvette into the holder and close the lid.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
10 mL
×
5