SensoControl

®

 

Operating Instruction 

ServiceJunior

 

 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

Please read carefully before use! 

 
 

 

 

 

 
 

 

 

Содержание SensoControl ServiceJunior

Страница 1: ...r Hannifin GmbH Co KG Tube Fittings Division Europe Am Metallwerk 9 33659 Bielefeld Phone 49 521 4048 0 Fax 49 521 4048 4280 SensoControl Operating Instruction ServiceJunior Please read carefully before use ...

Страница 2: ... hydraulics 6 5 Operating the ServiceJunior 7 5 1 Turning on ON 7 5 2 Turn off OFF 8 5 3 Turn on back light 8 5 4 MIN MAX Display 8 5 5 FS Full Scale Display 8 5 6 Erasing the MIN MAX values 8 5 7 OFL Display 8 5 8 Zero Point correction ZERO 9 5 9 Resetting the zero point correction 9 5 10 Automatic Power Off 10 5 11 Changing the Unit 11 5 12 Filter Settings 12 5 13 Display Serial Number 13 5 14 C...

Страница 3: ...per limit of the measurement range as well as the permissible temperature range must be observed For repairs or calibration of the measurement instruments please contact a Parker sales branch 1 2 Device versions and range of delivery Device versions Range of delivery Basic setting to unit bar Pressure port BSPP pressure gauge Delivery includes rubber protection 2 x batteries 1 5 VDC Mignon Range 1...

Страница 4: ...tteries When in continuous operation without light the service life of the batteries is 1 500 hours A battery symbol permanently displays the actual battery status Caution Turn off the device before replacing the batteries Open the battery compartment Insert the new batteries as depicted Ensure correct polarity of the batteries Batteries 2 x 1 5 V LR6 AA ...

Страница 5: ...400 bar MENU ZERO MENU ZERO Press for 2 sec Changes the unit Auto Power Off on off Zero point calibration RESET OK RESET OK Erases MIN and MAX values from the memory Confirms the MENU functions Display 4 digit LCD with back light function Displays measurement values and menu functions bar graph with peak hold function Actual value display 15 mm MIN MAX or Full Scale FS 8 mm Battery status Analogue...

Страница 6: ...y means of a pixel graduation mark The indicated value is refreshed at intervals of 50 ms 20 measurements sec ACT Indicates the actual pressure The indicated value is refreshed at intervals of 300 ms 3 times sec MIN MAX Indicates the MIN MAX or FS value according to setting The indicated value is refreshed at intervals of 300 ms 3 times sec FS Upper limit of the scale e g 400 bar Units Indicates t...

Страница 7: ...ction of engineering units bar mbar PSI kPa MPa By pressing the MENU key 2 s the desired function appears Skip to next function by pressing MENU again Confirmation by pressing the OK key The ServiceJunior returns into the display mode RESET OK RESET OK RESET OK RESET OK RESET OK MENU ZERO MENU ZERO MENU ZERO MENU ZERO MENU ZERO ...

Страница 8: ...n with the recommended sealing type the pressure port is compatible with following fluids media Fluid Media Sealing Diesel FKM Viton Gasoline FKM Viton Breakfluid EPDM Hydraulic Oil FKM Viton Air 4bar FKM Viton Observe specified torques when fitting the ServiceJunior The hex size of the pressure connection is 27 mm Pressure port Torque BSPP pressure gauge 35 Nm When fitting directly please ensure ...

Страница 9: ...dication of full scale range FS Unit bar SCJN xxx 01 Unit PSI SCJR xxx 02 MIN FS MAX REC x10 FS Auto Power Off function is active Power Off activates automatically switch off after 5 min This function can be altered in MENU MAX ON OFF Press Display mode ACT value displayed MAX peak ...

Страница 10: ...sure tests one after another the MIN MAX memory should be deleted RESET after every test cycle MIN MAX FS MIN MAX and FS value is indicated in the display 5 5 FS Full Scale Display Displaying the upper limit of the scale FS is designed to increase readability of the bar graph function The upper limit of the measurement range is indicated numerically FS is indicated in sequence after MIN and MAX MI...

Страница 11: ...asurements ensure no system pressure is being applied when carrying out this function Zero point correction cannot be carried out Please ensure that no system pressure is being applied This initiates the zero point correction The ACT actual value is indicated in the display as 0 0 bar The correction remains active until the device is turned off OFL ZEro are displayed for 3 seconds if the measured ...

Страница 12: ...le The settings Auto Power Off or Continuous operations remain stored and are active when the device is turned off and on again PO On When RESET OK is pressed the Auto Power Off is enabled The device will switch off after 5 min PO OFF When RESET OK is pressed the device must be turned off manually Auto Power Off Continuous operations ...

Страница 13: ...page 11 of 18 5 11 Changing the Unit MENU ZERO Press for 2 s MENU ZERO Press to skip MENU ZERO Press once briefly The next unit is indicated RESET OK Confirm unit selection ...

Страница 14: ...ction will smooth the actual read out shown on the display Following setup can be done Initially the actual setting will be displayed MENU ZERO Press once briefly Filter selection is indicated RESET OK Confirm filter configuration 0 No Filter function 1 Filter Level 1 2 Filter Level 2 3 Filter Level 3 ...

Страница 15: ...page 13 of 18 5 13 Display Serial Number MENU ZERO Press 2sec MENU ZERO Press MENU ZERO Press MENU ZERO Press RESET OK Press Display of serial number 1 line Display of software version 2 line ...

Страница 16: ...erviceJunior is connected to the VAS tester the actual display will be switched off The bar graph indicator is still in operation The described filter function 5 11 is not active now and need to be set up at the VAS tester device Following display appears ...

Страница 17: ... 4 0 60 0 400 bar Resolution 12 bit 4 096 steps Housing Ø 79 mm T 33 mm Zinc Die Cast with Rubber Protection TPE Weight 540 g Parts in contact with media Sensor Cell Ceramic Viton 1 1 bar 1 4 bar Stainless Steel 1 4534 0 60 0 400 bar Suitable fluids media ATF and Hydraulic oils brake fluid patrol diesel water and air 4 bar no oxygene Please use suitable sealings Power supply Battery 2 x1 5 VDC LR6...

Страница 18: ...bar 0 1 bar 0 3 bar 0 6 bar 0 400 0 bar 800 bar 1 700 bar 0 1 bar 2 0 bar 4 0 bar The ServiceJunior meets the guidelines of the European Community EU It is confirmed that this product is approved acc to following standards Technical subject to change February 2007 DIN EN 61000 6 2 DIN EN 61000 6 3 Burst pressures related to tests without assembled adapters Exceeding the maximum overload values Pma...