181
SIL
VER
#2
• Insert the second cuvette (#2) into the instrument.
• Press
Read
to start the reading. The instrument displays the results in
mg/L of silver (Ag).
INTERFERENCES
Interference may be caused by:
Al
3+
above 30 mg/L
Fe
2+
above 1.5 mg/L
Ca
2+
above 1000 mg/L (as CaCO
3
) Fe
3
+
above 10 mg/L
Cd
2+
above 20 mg/L
K
+
above 500 mg/L
Cl¯ above 8000 mg/L
Mn
2+
above 25 mg/L
Co
2+
above 1.5 mg/L
Mg
2+
above 1000 mg/L (as CaCO
3
)
Cr
3+
above 20 mg/L
Na
+
above 5000 mg/L
Cr
6+
above 40 mg/L
Ni
2+
above 1.5 mg/L
Cu
2+
above 15 mg/L
Pb
2+
above 20 mg/L
F¯ above 20 mg/L
Zn
2+
above 30 mg/L
• Add 3 drops of
HI93703‑51
Dispersing Agent, replace
the cap and invert gently for 10 seconds.