
158
OZONE
• Remove the cuvette.
• Add one packet of the
HI93703‑52‑0
Glycine Powder.
Replace the cap and shake gently until completely
dissolved.
• Add one packet of
HI93757‑0
Ozone Reagent. Replace
the cap and shake gently for 20 seconds.
• Replace the cuvette into the instrument and close the lid.
• Press
Timer
and the display will show the countdown prior to the measurement or, alternatively,
wait for 2 minutes and press
Read.
When the timer ends the meter will perform the reading.
Record this value as B.
• Press the
Zero
key. The display will show “‑0.0‑” the meter is zeroed and ready for measure.
#2
#1