127
NICKEL LOW RANGE
• Press
Timer
and the display will show a countdown or alternatively, wait for 15 minutes.
•Add one packet of
HI93740C‑0
Nickel Low Range Reagent C
to each beaker, cap and swirl to mix until completely dissolved.
•Fill one cuvette (#1) with 10 mL of the blank (up to the mark).
• Place the cuvette into the holder and close the lid.
• Press the
Zero
key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
• Fill a second cuvette (#2) with 10 mL of the reacted sample
(up to the mark).
• Insert the second cuvette into the instrument and close the lid.
10 mL
10 mL
#1
#2
10 mL