background image

12

13

• It is important that the sample does not contain any debris.

This would corrupt the reading.

• In order to avoid reagent leaking and to obtain more

accurate measurements, it is recommended to close the
cuvette first with the supplied HPDE plastic stopper and
then with the black cap.

• Each time the cuvette is used, the cap must be tightened

to the same degree.

• Whenever the cuvette is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with 

HI 731318

or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain

accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvette for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the 

validation procedure

 to

be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

MEASUREMENT PROCEDURE

STARTUP

Prepare the instrument for measurement as follows:
• Unpack the instrument by removing the dust protection sleeve from the instrument cuvette

holder.

• Place the battery in the instrument as described in the “BATTERY REPLACEMENT” chapter.
• Place the instrument on a flat table.
• Do not place the instrument under direct sun light.

To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking 

“ZERO”

 indicates that the

instrument needs to be zeroed first.

• Fill the cuvette with 10 mL of unreacted

sample, up to the mark, and replace the
cap.

• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Turn the meter on by pressing ON/OFF

ON/OFF

ON/OFF

ON/OFF

ON/OFF.

The display briefly shows all tags on.

Summary of Contents for HI 96715C

Page 1: ...1 Instruction Manual HI 96715C Ammonia Medium Range ISM 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Page 2: ...t carefully before using the meter All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA PRELIMINARY EXAMINATION Please examine this product carefully Make sure that the instrument is not damaged If any damage occurred during shipment please notify your Dealer Each HI 96715 I...

Page 3: ... an adaptation from ASTM Manual of Water and Environmental Technology D1426 93 Nessler Method The reagents are liquid form and are supplied in bottles The amount of reagent is precisely dosed to ensure the maximum repeatability degree Celsius degree Fahrenheit milligrams per liter mg L is equivalent to ppm parts per million milliliter Light Emitting Diode Liquid Crystal Display American Society fo...

Page 4: ...er is based on a Light Emitting Diode LED combined with a narrow band interference filter to guarantee both high performance and reliable results HI 96715 block diagram optical layout A microprocessor controlled LED emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette The optical path is fixed by the diameter of the cuvette Then the light is spectr...

Page 5: ...ion date or restore factory calibration DISPLAY ELEMENTS DESCRIPTION 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon shows the charging level of the battery 4 The hourglass appears when an internal checking is in progress 5 Status messages 6 The chronometer appears when the reaction timer i...

Page 6: ...g 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test c during calibration procedure Standar...

Page 7: ...on times reported in this manual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F In order to maximize accuracy prior to a measurement follow the validation procedure to be sure that the instrument is properly calibrated If necessary calibrate the instrument MEASUREMENT PROCEDURE STARTUP Prepare the instrument for measurement as follo...

Page 8: ...oth cases the lamp cuvette and detector icons will appear on the display depending on the measurement phase At the end of measurement the instrument directly displays concentration in mg L of ammonia nitrogen NH3 N on the LCD To convert the reading to mg L of ammonia NH3 multiply by a factor of 1 216 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibr...

Page 9: ...ications as reported in the CAL CHECKTM Standard Certificate If the value is found out of the specifications please check that the cuvettes are free of fingerprints oil or dirt and repeat validation If results are still found out of specifications then recalibrate the instrument Turn the meter on by pressing ON OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready Press...

Page 10: ...w seconds the display will show 0 0 The meter is now zeroed and ready for calibration The blinking READ asks for reading calibration standard Remove the cuvette Place the CAL CHECKTM Standard HI 96715 11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press READ TIMER and the lamp cuvette and detector icons will appear on the display de...

Page 11: ...ery indicator with three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line for 33 capacity Empty battery icon blinking if the capacity is under 10 If the battery is empty and accurate measurements can t be taken anymore the instrument shows dead batt dead batt dead batt dead batt dead batt and turns off To restart the instrument the battery must be replaced with a fresh one...

Page 12: ...e instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase Hanna Instrume...

Reviews: