background image

6

7

1) GLP/  key
2) CAL CHECK key
3) ZERO/CFM key
4) READ/ /TIMER key
5) ON/OFF key

6) Liquid Crystal Display (LCD)
7) Cuvette alignment indicator

8) Cuvette holder

PRINCIPLE OF OPERATION

FUNCTIONAL DESCRIPTION

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and
matter. When a light beam crosses a substance, some of the radiation may be absorbed by
atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length
through the matter and on the 

physical

-chemical characteristics of the substance according to the

Lambert-Beer Law:

-log

 

I

/

I

o

 = 

ε

λ

 

c d

or

A  

ε

λ

 

c d

Where:

-log

 

I

/

I

o

=

Absorbance (A)

I

o

=

intensity of incident light beam

I

=

intensity of light beam after absorption

ε

λ

=

molar extinction coefficient at wavelength

 

λ

c

=

molar concentration of the substance

d

=

optical path through the substance

Therefore, the concentration "c" can be calculated from the absorbance of the substance as the
other factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound
from a specific chemical reaction between sample and reagents. Given that the absorption of a
compound strictly depends on the wavelength of the incident light beam, a narrow spectral
bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna's 

HI 96715

  colorimeter is based on a Light Emitting Diode (LED)

combined with a narrow-band interference filter to guarantee both high performance and reliable
results.

HI 96715 block diagram (optical layout)

A microprocessor controlled LED emits radiation which is first optically conditioned and beamed
to the sample contained in the cuvette. The optical path is fixed by the diameter of the cuvette.
Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of
intensity 

I

o

 or 

I

.

The photoelectric cell collects the radiation 

I

 that is not absorbed by the sample and converts

it into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring
unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the
actual measurement is performed.
The cuvette has a very important role because it is an optical element and thus requires
particular attention. It is important that both, the measurement and the calibration (zeroing)
cuvettes, are optically identical to provide the same measurement conditions. Whenever possible
use the same cuvette for both. It is necessary that the surface of the cuvette is clean and not
scratched. This to avoid measurement interference due to unwanted reflection and absorption of
light. It is recommended not to touch the cuvette walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring
phases, it is necessary to close the cuvette to prevent any contamination.

INSTRUMENT DESCRIPTION

Summary of Contents for HI 96715C

Page 1: ...1 Instruction Manual HI 96715C Ammonia Medium Range ISM 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Page 2: ...t carefully before using the meter All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA PRELIMINARY EXAMINATION Please examine this product carefully Make sure that the instrument is not damaged If any damage occurred during shipment please notify your Dealer Each HI 96715 I...

Page 3: ... an adaptation from ASTM Manual of Water and Environmental Technology D1426 93 Nessler Method The reagents are liquid form and are supplied in bottles The amount of reagent is precisely dosed to ensure the maximum repeatability degree Celsius degree Fahrenheit milligrams per liter mg L is equivalent to ppm parts per million milliliter Light Emitting Diode Liquid Crystal Display American Society fo...

Page 4: ...er is based on a Light Emitting Diode LED combined with a narrow band interference filter to guarantee both high performance and reliable results HI 96715 block diagram optical layout A microprocessor controlled LED emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette The optical path is fixed by the diameter of the cuvette Then the light is spectr...

Page 5: ...ion date or restore factory calibration DISPLAY ELEMENTS DESCRIPTION 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon shows the charging level of the battery 4 The hourglass appears when an internal checking is in progress 5 Status messages 6 The chronometer appears when the reaction timer i...

Page 6: ...g 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test c during calibration procedure Standar...

Page 7: ...on times reported in this manual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F In order to maximize accuracy prior to a measurement follow the validation procedure to be sure that the instrument is properly calibrated If necessary calibrate the instrument MEASUREMENT PROCEDURE STARTUP Prepare the instrument for measurement as follo...

Page 8: ...oth cases the lamp cuvette and detector icons will appear on the display depending on the measurement phase At the end of measurement the instrument directly displays concentration in mg L of ammonia nitrogen NH3 N on the LCD To convert the reading to mg L of ammonia NH3 multiply by a factor of 1 216 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibr...

Page 9: ...ications as reported in the CAL CHECKTM Standard Certificate If the value is found out of the specifications please check that the cuvettes are free of fingerprints oil or dirt and repeat validation If results are still found out of specifications then recalibrate the instrument Turn the meter on by pressing ON OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready Press...

Page 10: ...w seconds the display will show 0 0 The meter is now zeroed and ready for calibration The blinking READ asks for reading calibration standard Remove the cuvette Place the CAL CHECKTM Standard HI 96715 11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press READ TIMER and the lamp cuvette and detector icons will appear on the display de...

Page 11: ...ery indicator with three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line for 33 capacity Empty battery icon blinking if the capacity is under 10 If the battery is empty and accurate measurements can t be taken anymore the instrument shows dead batt dead batt dead batt dead batt dead batt and turns off To restart the instrument the battery must be replaced with a fresh one...

Page 12: ...e instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase Hanna Instrume...

Reviews: