ERRORS AND WARNINGS

KEYPAD DESCRIPTION

ON/OFF

: to turn the meter on and off.

ZERO/CFM

: this is a bi-functional key. Just press to zero the meter prior to measurement,

or to confirm edited values. In  or to confirm factory calibration restore.

READ/ /TIMER

: this is a multi-functional key. In 

measurement mode, press to make a

measurement, or press and hold for three seconds to start a pre-programmed countdown prior
to measurement. In 

GLP mode press to view the next screen.

CAL CHECK

: this is a bi-functional key. Just press to perform the validation of the meter, or

press and hold for three seconds to enter 

calibration mode.

GLP

/

this is a bi-functional key. Just press to enter 

GLP mode. In calibration mode press

to edit the date and time.

OPERATING MODES

Measurement mode: default operation mode, enables both 

validation 

and 

measurement

.

Calibration mode: may be entered by keeping 

CAL CHECK

 pressed 

for three seconds (

the

“CAL”

 tag appears), it enables 

calibration 

of the instrument.

GLP mode may be entered by pressing 

GLP

/  (

“GLP”

 appears), it enables consulting of

user calibration date or restore factory calibration.

DISPLAY ELEMENTS DESCRIPTION

1) The measuring scheme (lamp, cuvette, detector), appears during different phases of zero or

reading measurement

2) Error messages and warnings
3) The battery icon shows the charging level of the battery
4) The hourglass appears when an internal checking is in progress

5) Status messages
6) The chronometer appears when the reaction timer is running
7) The month, day and date icons appear when a date is displayed

8) Four digit main display

9) Measuring units

10) Four digit secondary display

The instrument shows clear messages when erroneous condition appears. Messages are also displayed
when the obtained values are outside expected range. The beeper is playing a beep on errors.

a) on zero reading

Light High:

 There is too much light to perform a measurement.

Please check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement.

Please check the preparation of the zero cuvette.

No Light:

The instrument cannot adjust the light level. Please

check that the sample does not contain any debris.

Inverted cuvettes:

 The sample and the zero cuvette are inverted.

Zero:

 A zero reading was not taken. Follow the instructions of

the measurement procedure for zeroing the meter.

b) on sample reading

Summary of Contents for HI 96715C

Page 1: ...1 Instruction Manual HI 96715C Ammonia Medium Range ISM 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Page 2: ...t carefully before using the meter All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA PRELIMINARY EXAMINATION Please examine this product carefully Make sure that the instrument is not damaged If any damage occurred during shipment please notify your Dealer Each HI 96715 I...

Page 3: ... an adaptation from ASTM Manual of Water and Environmental Technology D1426 93 Nessler Method The reagents are liquid form and are supplied in bottles The amount of reagent is precisely dosed to ensure the maximum repeatability degree Celsius degree Fahrenheit milligrams per liter mg L is equivalent to ppm parts per million milliliter Light Emitting Diode Liquid Crystal Display American Society fo...

Page 4: ...er is based on a Light Emitting Diode LED combined with a narrow band interference filter to guarantee both high performance and reliable results HI 96715 block diagram optical layout A microprocessor controlled LED emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette The optical path is fixed by the diameter of the cuvette Then the light is spectr...

Page 5: ...ion date or restore factory calibration DISPLAY ELEMENTS DESCRIPTION 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon shows the charging level of the battery 4 The hourglass appears when an internal checking is in progress 5 Status messages 6 The chronometer appears when the reaction timer i...

Page 6: ...g 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test c during calibration procedure Standar...

Page 7: ...on times reported in this manual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F In order to maximize accuracy prior to a measurement follow the validation procedure to be sure that the instrument is properly calibrated If necessary calibrate the instrument MEASUREMENT PROCEDURE STARTUP Prepare the instrument for measurement as follo...

Page 8: ...oth cases the lamp cuvette and detector icons will appear on the display depending on the measurement phase At the end of measurement the instrument directly displays concentration in mg L of ammonia nitrogen NH3 N on the LCD To convert the reading to mg L of ammonia NH3 multiply by a factor of 1 216 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibr...

Page 9: ...ications as reported in the CAL CHECKTM Standard Certificate If the value is found out of the specifications please check that the cuvettes are free of fingerprints oil or dirt and repeat validation If results are still found out of specifications then recalibrate the instrument Turn the meter on by pressing ON OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready Press...

Page 10: ...w seconds the display will show 0 0 The meter is now zeroed and ready for calibration The blinking READ asks for reading calibration standard Remove the cuvette Place the CAL CHECKTM Standard HI 96715 11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press READ TIMER and the lamp cuvette and detector icons will appear on the display de...

Page 11: ...ery indicator with three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line for 33 capacity Empty battery icon blinking if the capacity is under 10 If the battery is empty and accurate measurements can t be taken anymore the instrument shows dead batt dead batt dead batt dead batt dead batt and turns off To restart the instrument the battery must be replaced with a fresh one...

Page 12: ...e instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase Hanna Instrume...