background image

10

• Press the CFM key to confirm the calibra-

tion: the meter stores the first calibration

point; the primary LCD will show the

calibrated reading, while the secondary

LCD will show the second buffer to be

used for calibration (pH 4.01).

If you’re going to calibrate with a

different buffer, select the desired value

by pressing the arrow keys.

• Submerge the electrode approximately 4 cm (1½") into the

second buffer solution, place the temperature probe as close as

possible to the electrode and stir gently.

• The LCD will flash "NOT READY" for 12 seconds, then:

if the reading is not close to the selected buffer, "WRONG 

 

"

and "WRONG  " will blink alternatively;

if it is close to the selected buffer the meter

will advise the user with an acoustic

signal (if enabled) when the reading

becomes stable and the display will change

to "READY" and blinking "CFM".

• Press the CFM key: the value is stored in memory and the meter

returns to normal mode. The tags corresponding to the buffers

used for calibration will light up together with the “condition”

bargraph (if enabled).

Note:

The meter automatically skips the buffer used for the first calibration

point to avoid erroneous procedure. A difference of at least 1.5 pH

unit is required between the two buffers used for the offset and slope

calibration: once calibrated at either pH 7.01 or 6.86, the

instrument automatically ignores the other value for the second point

(same for pH 10.01 and 9.18).

Note:

During calibration, the secondary LCD shows the selected buffer

value; press RANGE to display the buffer temperature.

Note:

To clear a previous calibration and return to the default

values, press CLR at any time after entering the calibration

mode. The LCD will show “CLr CAL” for one second, and then

will return to normal measurement mode.

The LCD will show an empty bargraph to warn the user that

the meter is not calibrated.

19

For refillable electrodes, if the refill solution (electrolyte) is more than 2½

cm (1") below the fill hole, add the appropriate Electrolyte Solution.

MEASUREMENT

Rinse the electrode tip with distilled water, immerse it (4 cm / 1½")

in the sample and stir gently for a few seconds.

For a faster response and to avoid cross contamination of the samples,

rinse the electrode tip with the solution to be tested, before taking

any measurements.

STORAGE PROCEDURE

To minimize clogging and ensure a quick response time, the glass

bulb and the junction should always be kept moist.

When not in use, store it with a few drops of 

HI 70300 

storage

solution in the protective cap.

NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED

WATER.

PERIODIC MAINTENANCE

Inspect electrode and cable. The cable used for the connection to the

meter must be intact and there must be no points of broken

insulation on the cable or cracks on the electrode stem or bulb. If any

scratches or cracks are present, replace the electrode. Rinse off any salt

deposits with water.

Connectors must be perfectly clean and dry.

For refillable electrodes:

Refill the electrode with fresh electrolyte (see the electrode’s specifications

to select the correct refilling solution). Allow the electrode to stand

upright for 1 hour. Follow the Storage Procedure above.

CLEANING PROCEDURE

• 

General

Soak in Hanna 

HI 7061

 General Cleaning

Solution for approximately ½ hour.

• 

Protein

Soak in Hanna 

HI 7073

 Protein Cleaning

Solution for 15 min.

• 

Inorganic

Soak in Hanna 

HI 7074

 Inorganic

CleaningSolution for 15 min.

• 

Oil/grease

Rinse with Hanna 

HI 7077

 Oil & Fat Cleaning

Solution for 1 min.

IMPORTANT:

 After performing any of the cleaning procedures, rinse

the electrode thoroughly with distilled water and soak it in 

HI 70300

Storage Solution for at least 1 hour before taking measurements.

Summary of Contents for HI 9026

Page 1: ...el 21 4584 2941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 662 619 0708 11 Fax 662 619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765...

Page 2: ...ts prepaid When shipping any instrument make sure it is properly packed for complete protection TABLE OF CONTENTS WARRANTY 2 PRELIMINARY EXAMINATION 3 GENERALDESCRIPTION 3 FUNCTIONAL DESCRIPTION 4 SPECIFICATIONS 5 OPERATIONAL GUIDE 6 pH CALIBRATION 9 pHBUFFERTEMPERATUREDEPENDENCE 13 SETUP MENU 14 mV CALIBRATION 15 TEMPERATURE CALIBRATION 15 BATTERYREPLACEMENT 15 LCDMESSAGEGUIDE 16 TEMPERATURE CORR...

Page 3: ...irely new dimension to the measuremeny of pH by allowing the user to dramatically improve the reliability of the measurement 7 memorized buffers pH 1 68 4 01 6 86 7 01 9 18 10 01 and 12 45 for calibration Custom calibration by entering up to two custom buffers Messages on the LCD to make the calibration easy and accurate Diagnostic features to alert the user when the electrode needs cleaning Monit...

Page 4: ... 13 Primary display FUNCTIONAL DESCRIPTION 25 Kynar is registered Trademark of Pennwalt Corp Ultem is registered Trademark of General Electric Co HI 1413B Glass body single junction flat tip Viscolene non refillable combination pH electrode Use surface measurement ORP ELECTRODES HI 3131B Glass body refillable combination platinum ORP electrode Use titration HI 3230B Plastic body Ultem gel filled c...

Page 5: ...y r e t t a B s e i r e t t a b e n i l a k l a e z i s A A V 5 1 x 4 e s u s u o u n i t n o c f o s r u o h 0 0 5 x o r p p a f f o o t u A r o s e t u n i m 0 2 e l b a t c e l e s r e s U d e l b a s i d s n o i s n e m i D 4 2 x 1 3 x 7 7 m m 0 6 x 0 8 x 6 9 1 y l n o r e t e m t h g i e W z o 5 1 g 5 2 4 t n e m n o r i v n E F 2 2 1 2 3 C 0 5 0 0 0 1 H R x a m y t n a r r a W s r a e y 2 24...

Page 6: ...pressed to save battery life To disable this feature see Setup Menu section on page 9 pH MEASUREMENTS To take a pH measurement remove the elec trode protective cap and simply submerge the tip 4cm 1 of the electrode and the temperature probe into the sample to be tested If necessary press the RANGE key until the display changes to the pH mode Allow for the electrode to adjust and reading to stabili...

Page 7: ...MEASUREMENTS To perform ORP measurements connect an optional ORP electrode see Accessories section to the meter and turn it ON If necessary enter the mV mode by pressing RANGE until the display changes to mV Submerge the ORP electrode tip 4cm 1 into the sample to be tested and wait a few minutes for the reading to stabilize Measurements within the 699 9 mV range are displayed with 0 1 mV resolutio...

Page 8: ...olution 20 mL sachet 25 pcs HI 70010P pH 10 01 Buffer Solution 20 mL sachet 25 pcs HI 7001L pH 1 68 Buffer Solution 500 mL bottle HI 7001M pH 1 68 Buffer Solution 230 mL bottle HI 7004L pH 4 01 Buffer Solution 500 mL bottle HI 7004M pH 4 01 Buffer Solution 230 mL bottle HI 7006L pH 6 86 Buffer Solution 500 mL bottle HI 7006M pH 6 86 Buffer Solution 230 mL bottle HI 7007M pH 7 01 Buffer Solution 50...

Page 9: ...l e w o L s e d o r t c e l e e l b a l l i f e r y l n o e d o r t c e l e e h t n a e l C e t y l o r t c e l e h s e r f h t i w l l i f e R y l n o s e d o r t c e l e e l b a l l i f e r s w o h s y a l p s i D e l a c s l l u f g n i k n i l b e u l a v f o t u o g n i d a e R e g n a r s i e l p m a s t a h t k c e h C e g n a r e l b a r u s a e m n i h t i w d n a l e v e l e t y l o r t ...

Page 10: ... to normal measurement mode The LCD will show an empty bargraph to warn the user that the meter is not calibrated 19 For refillable electrodes if the refill solution electrolyte is more than 2 cm 1 below the fill hole add the appropriate Electrolyte Solution MEASUREMENT Rinse the electrode tip with distilled water immerse it 4 cm 1 in the sample and stir gently for a few seconds For a faster respo...

Page 11: ...trument is provided with a real time clock RTC in order to monitor the time elapsed since the last pH calibration The real time clock is reset every time the meter is calibrated and the expired calibration status is triggered when the meter detects a calibration time out The CAL tag will start blinking to warn the user that the meter should be recalibrated The calibration time out can be set see S...

Page 12: ...are cleared the comparison is done with the default values 17 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4 months 120 C Less than 1 month Alkaline Error High concentrations of sodium ions interfere with readings in alkaline solutions The pH at which the interference starts to be significant depends upon the composition of the glass This interference is called alkaline erro...

Page 13: ... 0 0 1 1 5 8 5 8 1 8 7 1 7 1 4 7 8 6 2 0 7 7 8 8 4 7 9 1 9 0 1 0 9 4 9 1 9 7 1 9 1 4 8 8 6 3 0 7 5 8 8 5 7 9 2 8 0 1 5 9 3 0 2 1 8 1 0 2 4 9 8 6 4 0 7 3 8 8 6 7 9 3 7 0 1 16 LCD MESSAGE GUIDE TAGS SYMBOLS Mode tags light up for indicating the corresponding active mode and blink for warning the user SETUP on setup menu mode has been entered CFM blinking ask confirmation of calibration value CAL on ...

Page 14: ...batteries soon The meter is also provided with the BEPS BatteryErrorPreventionSystem feature which automatically turns the instrument off when the battery level is too low to ensure reliable readings At start up the display will show 0 bAtt for a few seconds then the meter automatically turns off Battery replacement must only take place in a non hazardous area using the battery types specified in ...

Reviews: