background image

12

Notes:

• When the meter is not calibrated or calibration is cleared (default

values loaded) there is no “expired calibration”, and the display

always shows a blinking CAL tag.

• When an abnormal condition in the RTC is detected the meter

forces the “expired calibration” status.

CONDITION

The display is provided with a 5-dot bargraph (unless the feature is

disabled) which gives an indication of the electrode status after

calibration as follows:

Bargraph indication

Condition value

All 5 dots steady

81 to 100% of life

4 dots steady

61 to 80%

3 dots steady

41 to 60%

2 dots steady

21 to 40%

1 dot steady

1 to 20%

1 dot blinking

0%

Only frame is ON

No info available

The “condition” bargraph remains active for 12 hours after calibration,

then only the frame is shown.

Note: 

When an abnormal condition in the RTC is detected, the “condition”

is cleared and only the bargraph frame is shown on the display.

CLEAN ELECTRODE

Each time pH calibration is performed, the meter internally compares

the new calibration with the one previously stored.

When this comparison indicates a significant

difference, the CLEAN message blinks on the

LCD to advise the user that the pH electrode

may need to be cleaned (see “Electrode

Conditioning & Maintenance” section).

After cleaning, perform calibration.

Note: 

If the calibration data are cleared, the comparison is done

with the default values.

17

Typical Electrode Life

Ambient Temperature

1 – 3 years

90 °C

Less than 4 months

120 °C

Less than 1 month

Alkaline Error

High concentrations of sodium ions interfere with readings in alkaline

solutions. The pH at which the interference starts to be significant

depends upon the composition of the glass. This interference is called

alkaline error and causes the pH to be underestimated. Hanna’s glass

formulations have the indicated characteristics.

1.0 Mol L

-1

 Na

+

0.1 Mol L

-1

 Na

+

Sodium Ion Correction for Glass at 20

-

25 

°

C

Concentration

pH

  Error

13.00

13.50

14.00

12.50

13.00

13.50

14.00

0.10

0.14

0.20

0.10

0.18

0.29

0.40

The resistance of glass electrodes partially depends on the temperature.

The lower the temperature, the higher the resistance. It takes more time

for the reading to stabilize if the resistance is higher. In addition, the

response time will suffer to a greater degree at temperatures below 25 °C.

Since the resistance of the pH electrode is in the range of 50 – 200

Mohms, the current across the membrane is in the pico Ampere range.

Large currents can disturb the calibration of the electrode for many

hours.

For these reasons high humidity environments, short circuits and static

discharges can be detrimental to a stable pH reading.

The pH electrode’s life also depends on the temperature. If constantly

used at high temperatures, the electrode life is drastically reduced.

TEMPERATURE CORRELATION

FOR pH SENSITIVE GLASS

Summary of Contents for HI 9026

Page 1: ...el 21 4584 2941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 662 619 0708 11 Fax 662 619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765...

Page 2: ...ts prepaid When shipping any instrument make sure it is properly packed for complete protection TABLE OF CONTENTS WARRANTY 2 PRELIMINARY EXAMINATION 3 GENERALDESCRIPTION 3 FUNCTIONAL DESCRIPTION 4 SPECIFICATIONS 5 OPERATIONAL GUIDE 6 pH CALIBRATION 9 pHBUFFERTEMPERATUREDEPENDENCE 13 SETUP MENU 14 mV CALIBRATION 15 TEMPERATURE CALIBRATION 15 BATTERYREPLACEMENT 15 LCDMESSAGEGUIDE 16 TEMPERATURE CORR...

Page 3: ...irely new dimension to the measuremeny of pH by allowing the user to dramatically improve the reliability of the measurement 7 memorized buffers pH 1 68 4 01 6 86 7 01 9 18 10 01 and 12 45 for calibration Custom calibration by entering up to two custom buffers Messages on the LCD to make the calibration easy and accurate Diagnostic features to alert the user when the electrode needs cleaning Monit...

Page 4: ... 13 Primary display FUNCTIONAL DESCRIPTION 25 Kynar is registered Trademark of Pennwalt Corp Ultem is registered Trademark of General Electric Co HI 1413B Glass body single junction flat tip Viscolene non refillable combination pH electrode Use surface measurement ORP ELECTRODES HI 3131B Glass body refillable combination platinum ORP electrode Use titration HI 3230B Plastic body Ultem gel filled c...

Page 5: ...y r e t t a B s e i r e t t a b e n i l a k l a e z i s A A V 5 1 x 4 e s u s u o u n i t n o c f o s r u o h 0 0 5 x o r p p a f f o o t u A r o s e t u n i m 0 2 e l b a t c e l e s r e s U d e l b a s i d s n o i s n e m i D 4 2 x 1 3 x 7 7 m m 0 6 x 0 8 x 6 9 1 y l n o r e t e m t h g i e W z o 5 1 g 5 2 4 t n e m n o r i v n E F 2 2 1 2 3 C 0 5 0 0 0 1 H R x a m y t n a r r a W s r a e y 2 24...

Page 6: ...pressed to save battery life To disable this feature see Setup Menu section on page 9 pH MEASUREMENTS To take a pH measurement remove the elec trode protective cap and simply submerge the tip 4cm 1 of the electrode and the temperature probe into the sample to be tested If necessary press the RANGE key until the display changes to the pH mode Allow for the electrode to adjust and reading to stabili...

Page 7: ...MEASUREMENTS To perform ORP measurements connect an optional ORP electrode see Accessories section to the meter and turn it ON If necessary enter the mV mode by pressing RANGE until the display changes to mV Submerge the ORP electrode tip 4cm 1 into the sample to be tested and wait a few minutes for the reading to stabilize Measurements within the 699 9 mV range are displayed with 0 1 mV resolutio...

Page 8: ...olution 20 mL sachet 25 pcs HI 70010P pH 10 01 Buffer Solution 20 mL sachet 25 pcs HI 7001L pH 1 68 Buffer Solution 500 mL bottle HI 7001M pH 1 68 Buffer Solution 230 mL bottle HI 7004L pH 4 01 Buffer Solution 500 mL bottle HI 7004M pH 4 01 Buffer Solution 230 mL bottle HI 7006L pH 6 86 Buffer Solution 500 mL bottle HI 7006M pH 6 86 Buffer Solution 230 mL bottle HI 7007M pH 7 01 Buffer Solution 50...

Page 9: ...l e w o L s e d o r t c e l e e l b a l l i f e r y l n o e d o r t c e l e e h t n a e l C e t y l o r t c e l e h s e r f h t i w l l i f e R y l n o s e d o r t c e l e e l b a l l i f e r s w o h s y a l p s i D e l a c s l l u f g n i k n i l b e u l a v f o t u o g n i d a e R e g n a r s i e l p m a s t a h t k c e h C e g n a r e l b a r u s a e m n i h t i w d n a l e v e l e t y l o r t ...

Page 10: ... to normal measurement mode The LCD will show an empty bargraph to warn the user that the meter is not calibrated 19 For refillable electrodes if the refill solution electrolyte is more than 2 cm 1 below the fill hole add the appropriate Electrolyte Solution MEASUREMENT Rinse the electrode tip with distilled water immerse it 4 cm 1 in the sample and stir gently for a few seconds For a faster respo...

Page 11: ...trument is provided with a real time clock RTC in order to monitor the time elapsed since the last pH calibration The real time clock is reset every time the meter is calibrated and the expired calibration status is triggered when the meter detects a calibration time out The CAL tag will start blinking to warn the user that the meter should be recalibrated The calibration time out can be set see S...

Page 12: ...are cleared the comparison is done with the default values 17 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4 months 120 C Less than 1 month Alkaline Error High concentrations of sodium ions interfere with readings in alkaline solutions The pH at which the interference starts to be significant depends upon the composition of the glass This interference is called alkaline erro...

Page 13: ... 0 0 1 1 5 8 5 8 1 8 7 1 7 1 4 7 8 6 2 0 7 7 8 8 4 7 9 1 9 0 1 0 9 4 9 1 9 7 1 9 1 4 8 8 6 3 0 7 5 8 8 5 7 9 2 8 0 1 5 9 3 0 2 1 8 1 0 2 4 9 8 6 4 0 7 3 8 8 6 7 9 3 7 0 1 16 LCD MESSAGE GUIDE TAGS SYMBOLS Mode tags light up for indicating the corresponding active mode and blink for warning the user SETUP on setup menu mode has been entered CFM blinking ask confirmation of calibration value CAL on ...

Page 14: ...batteries soon The meter is also provided with the BEPS BatteryErrorPreventionSystem feature which automatically turns the instrument off when the battery level is too low to ensure reliable readings At start up the display will show 0 bAtt for a few seconds then the meter automatically turns off Battery replacement must only take place in a non hazardous area using the battery types specified in ...

Reviews: