Hanna Instruments HI 9026 Instruction Manual Download Page 11

11

ONE-POINT CALIBRATION

For optimum accuracy it is always recommended to perform a two-

point calibration, but for a faster operation it is also possible to carry

out a single-point procedure.

Buffers at pH 7.01 or pH 6.86 (NIST) are normally the most

appropriate for this purpose, even though the meter can be calibrated

with any of the memorized calibration values.

After calibrating the first point (see above), press the CAL key to end

the calibration procedure.

Note:

With one-point calibration there is no “Condition” and only the

frame is shown. Calibration time-out is active.

CUSTOM BUFFER SELECTION

• The meter also allows the user to calibrate with custom buffers.

• To select a custom buffer press the CUSTOM BUF key while in calibration

mode. Use the arrow keys to set the desired value for C1 (the default

value is 7.00).

Note:

To increase the speed, keep pressed the UP or DOWN key.

• Once the desired value is reached, the meter will continue

calibration as with memorized buffers.

• Pressing again CUSTOM BUF, the meter

will select the second custom buffer. Pressing

CUSTOM BUF once again, the meter will

return to the standard buffers mode.

EXPIRED CALIBRATION

The instrument is provided with a real time clock (RTC), in order to

monitor the time elapsed since the last pH calibration.

The real time clock is reset every time the meter is calibrated and the

“expired calibration” status is triggered when the meter detects a

calibration time-out. The CAL tag will start blinking to warn the user

that the meter should be recalibrated.

The calibration time-out can be set (see “Setup menu” section on

page 14) from 0 (function disabled) to 14 days.

For example, if a 4 days time-out has been selected the meter will

issue the alarm exactly 4 days after the last calibration.

However, if at any moment the expiration value is changed (e.g. to

7 days), then the alarm will be immediately recalculated and appear

7 days after the last calibration.

18

PREPARATION PROCEDURE

Remove the electrode protective cap.

DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is

normal with electrodes and they will disappear when rinsed with water.

During transport tiny bubbles of air may have formed inside the glass

bulb. The electrode cannot function properly under these conditions.

These bubbles can be removed by "shaking down" the electrode as you

would do with a glass thermometer.

If the bulb and/or junction are dry, soak the electrode in 

HI 70300

Storage Solution for at least one hour.

ELECTRODE CONDITIONING & MAINTENANCE

 Not present in gel electrodes.

Summary of Contents for HI 9026

Page 1: ...el 21 4584 2941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 662 619 0708 11 Fax 662 619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765...

Page 2: ...ts prepaid When shipping any instrument make sure it is properly packed for complete protection TABLE OF CONTENTS WARRANTY 2 PRELIMINARY EXAMINATION 3 GENERALDESCRIPTION 3 FUNCTIONAL DESCRIPTION 4 SPECIFICATIONS 5 OPERATIONAL GUIDE 6 pH CALIBRATION 9 pHBUFFERTEMPERATUREDEPENDENCE 13 SETUP MENU 14 mV CALIBRATION 15 TEMPERATURE CALIBRATION 15 BATTERYREPLACEMENT 15 LCDMESSAGEGUIDE 16 TEMPERATURE CORR...

Page 3: ...irely new dimension to the measuremeny of pH by allowing the user to dramatically improve the reliability of the measurement 7 memorized buffers pH 1 68 4 01 6 86 7 01 9 18 10 01 and 12 45 for calibration Custom calibration by entering up to two custom buffers Messages on the LCD to make the calibration easy and accurate Diagnostic features to alert the user when the electrode needs cleaning Monit...

Page 4: ... 13 Primary display FUNCTIONAL DESCRIPTION 25 Kynar is registered Trademark of Pennwalt Corp Ultem is registered Trademark of General Electric Co HI 1413B Glass body single junction flat tip Viscolene non refillable combination pH electrode Use surface measurement ORP ELECTRODES HI 3131B Glass body refillable combination platinum ORP electrode Use titration HI 3230B Plastic body Ultem gel filled c...

Page 5: ...y r e t t a B s e i r e t t a b e n i l a k l a e z i s A A V 5 1 x 4 e s u s u o u n i t n o c f o s r u o h 0 0 5 x o r p p a f f o o t u A r o s e t u n i m 0 2 e l b a t c e l e s r e s U d e l b a s i d s n o i s n e m i D 4 2 x 1 3 x 7 7 m m 0 6 x 0 8 x 6 9 1 y l n o r e t e m t h g i e W z o 5 1 g 5 2 4 t n e m n o r i v n E F 2 2 1 2 3 C 0 5 0 0 0 1 H R x a m y t n a r r a W s r a e y 2 24...

Page 6: ...pressed to save battery life To disable this feature see Setup Menu section on page 9 pH MEASUREMENTS To take a pH measurement remove the elec trode protective cap and simply submerge the tip 4cm 1 of the electrode and the temperature probe into the sample to be tested If necessary press the RANGE key until the display changes to the pH mode Allow for the electrode to adjust and reading to stabili...

Page 7: ...MEASUREMENTS To perform ORP measurements connect an optional ORP electrode see Accessories section to the meter and turn it ON If necessary enter the mV mode by pressing RANGE until the display changes to mV Submerge the ORP electrode tip 4cm 1 into the sample to be tested and wait a few minutes for the reading to stabilize Measurements within the 699 9 mV range are displayed with 0 1 mV resolutio...

Page 8: ...olution 20 mL sachet 25 pcs HI 70010P pH 10 01 Buffer Solution 20 mL sachet 25 pcs HI 7001L pH 1 68 Buffer Solution 500 mL bottle HI 7001M pH 1 68 Buffer Solution 230 mL bottle HI 7004L pH 4 01 Buffer Solution 500 mL bottle HI 7004M pH 4 01 Buffer Solution 230 mL bottle HI 7006L pH 6 86 Buffer Solution 500 mL bottle HI 7006M pH 6 86 Buffer Solution 230 mL bottle HI 7007M pH 7 01 Buffer Solution 50...

Page 9: ...l e w o L s e d o r t c e l e e l b a l l i f e r y l n o e d o r t c e l e e h t n a e l C e t y l o r t c e l e h s e r f h t i w l l i f e R y l n o s e d o r t c e l e e l b a l l i f e r s w o h s y a l p s i D e l a c s l l u f g n i k n i l b e u l a v f o t u o g n i d a e R e g n a r s i e l p m a s t a h t k c e h C e g n a r e l b a r u s a e m n i h t i w d n a l e v e l e t y l o r t ...

Page 10: ... to normal measurement mode The LCD will show an empty bargraph to warn the user that the meter is not calibrated 19 For refillable electrodes if the refill solution electrolyte is more than 2 cm 1 below the fill hole add the appropriate Electrolyte Solution MEASUREMENT Rinse the electrode tip with distilled water immerse it 4 cm 1 in the sample and stir gently for a few seconds For a faster respo...

Page 11: ...trument is provided with a real time clock RTC in order to monitor the time elapsed since the last pH calibration The real time clock is reset every time the meter is calibrated and the expired calibration status is triggered when the meter detects a calibration time out The CAL tag will start blinking to warn the user that the meter should be recalibrated The calibration time out can be set see S...

Page 12: ...are cleared the comparison is done with the default values 17 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4 months 120 C Less than 1 month Alkaline Error High concentrations of sodium ions interfere with readings in alkaline solutions The pH at which the interference starts to be significant depends upon the composition of the glass This interference is called alkaline erro...

Page 13: ... 0 0 1 1 5 8 5 8 1 8 7 1 7 1 4 7 8 6 2 0 7 7 8 8 4 7 9 1 9 0 1 0 9 4 9 1 9 7 1 9 1 4 8 8 6 3 0 7 5 8 8 5 7 9 2 8 0 1 5 9 3 0 2 1 8 1 0 2 4 9 8 6 4 0 7 3 8 8 6 7 9 3 7 0 1 16 LCD MESSAGE GUIDE TAGS SYMBOLS Mode tags light up for indicating the corresponding active mode and blink for warning the user SETUP on setup menu mode has been entered CFM blinking ask confirmation of calibration value CAL on ...

Page 14: ...batteries soon The meter is also provided with the BEPS BatteryErrorPreventionSystem feature which automatically turns the instrument off when the battery level is too low to ensure reliable readings At start up the display will show 0 bAtt for a few seconds then the meter automatically turns off Battery replacement must only take place in a non hazardous area using the battery types specified in ...

Reviews: